Intramedullary Bone Saw Blade Assembly

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Intramedullary Bone Saw Blade Assembly
Issue: Medical devices - Sterility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Intramedullary Bone Saw Blade Assembly67818764475630
Intramedullary Bone Saw Blade Assembly67818770475635
Intramedullary Bone Saw Blade Assembly67818791475620

Issue

Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Sawblade Assembly due to potential sterile barrier packaging damaged caused by the exposed saw blade. It was identified internally that the tip protector was incorrectly placed on the blunt end of the device instead of the sawblade, leaving the saw blade exposed. The exposed saw blade has potential to damage the sterile barrier pouch, which has potential to result in loss of sterility. There have been zero complaints received.