Health Products

Astral 150™

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Astral 150™
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Astral 150™More than 10 numbers, contact manufacturer.27073
Astral 150™More than 10 numbers, contact manufacturer.27003

Issue

An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the ventilator's Printed Circuit Board Assembly ("PCBA"), causing the ventilator to inadvertently enter a fail-safe state. In this state, the ventilator is no longer able to deliver therapy. Alternative means of ventilation must be provided.