Health Products

Solar™ GI

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: Solar™ GI
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Solar™ GI24710569G3-8

Issue

Under extended clinical use conditions (e.g., long-duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter (part of MPP T/PC tubing set) and pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed. Water ingress in the air compressor can cause inaccurate pressure readings, resulting in excess fluid being administered to the patient than what is indicated by the system, which can lead to patient water intoxication. The issue is only related to Solar GI systems that have water perfusion capabilities.