Health Products

V-Loc™ 90 and 180 Suture Products

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: V-Loc™ 90 and 180 Suture Products
Issue: Medical devices - Sterility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
V-Loc 180 Wound Closure Device Needle P-12A6A0972VFYVLOCL0014
V-Loc 180 Abs Close Device GS-21A5M1060VYVLOCL0316
V-Loc 90 Absorbable Wound Closure Device P-12A5M1022VFYVLOCM0003
V-Loc 180 Absorbable Wound Closure Device - Green ColorA6A0998VYVLOCL2246
V-Loc 180 Absorbable Wound Closure Device - Green ColorA6A0996VYVLOCL0306
V-Loc 180 Wound Closure Device Needle V-20A5M1180VYVLOCL0604
V-Loc 180 Wound Closure Device Needle V-20A5M1181VYVLOCL0604
V-Loc 180 Wound Closure Device Needle CV-23A5M1064VYVLOCL0804

Issue

On 29-Jan-2026 during a quality inspection at a manufacturing site, a V-Loc 180 unit was found with a hole in the breather pouch on the Tyvek side. The breather pouch serves as the sterile barrier system for the suture. Investigation determined the briller plate on the sealing station may be causing the hole. No complaints have been identified for this issue, but lots within the issue bounding have been released to Distribution Centers (DCs) and sold to customers. These released lots may also contain units with holes in the Tyvek side of the breather pouch.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime