Health Products

Atellica® CH Urine Albumin (UAlb)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Atellica® CH Urine Albumin (UAlb)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Atellica® CH Urine Albumin (UAlb)More than 10 numbers, contact manufacturer.11537225

Issue

Siemens Healthineers has confirmed, through the investigation of customer complaints, the potential for the following issues associated with processing Atellica CH Urine Albumin (UAlb) tests on Atellica CH and Atellica CI analyzers.• Falsely depressed Atellica CH Urea Nitrogen (UN_c) results• Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results• Increased rate of reaction cuvette QC failuresThese issues may impact quality control (QC), calibrators, and patient results for the UN_c and A1c_E assays across all sample types.