· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Suture Cinch - Long | 37663465 | M00505550 |
| Suture Cinch - Long | 37086263 | M00505550 |
| Suture Cinch - Long | 37604205 | M00505550 |
| Suture Cinch - Long | 37073352 | M00505550 |
| Suture Cinch - Long | 37586915 | M00505550 |
| Suture Cinch - Long | 37302211 | M00505550 |
| OverStitch™ Suture Cinch | 38344889 | M00505540 |
| OverStitch™ Suture Cinch | 38324701 | M00505540 |
| OverStitch™ Suture Cinch | 37292127 | M00505540 |
| OverStitch™ Suture Cinch | 37003596 | M00505540 |
| OverStitch™ Suture Cinch | 38334198 | M00505540 |
| OverStitch™ Suture Cinch | 37940955 | M00505540 |
| OverStitch™ Suture Cinch | 37004909 | M00505540 |
| OverStitch™ Suture Cinch | 38337305 | M00505540 |
| OverStitch™ Suture Cinch | 37307313 | M00505540 |
| OverStitch™ Suture Cinch | 36938264 | M00505540 |
| OverStitch™ Suture Cinch | 37352684 | M00505540 |
| OverStitch™ Suture Cinch | 37688046 | M00505540 |
Boston Scientific is initiating removal of specific lots of OverStitch™ Suture Cinch and Suture Cinch-Long (collectively "Suture Cinch"). these devices are accessories used with the OverStitch™ Endoscopic Suturing System, OverStitch™ NXT Endoscopic Suturing System, and the X-Tack™ Endoscopic HeliX Tacking System to secure and cut the proximal end of a suture through the working channel of a gastroscope or colonoscope.During device activation, the Suture Cinch may fail to deploy the cinch implant and/or fail to cut the suture as intended. The issue cannot be detected through pre-use inspection but becomes apparent once the user attempts to deploy the device during the procedure.