Health ProductsType I recall

Sphere-9™ Catheter

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Sphere-9™ Catheter
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Sphere-9™ CatheterAll lots.AFR-00001

Issue

There have been forty-six (46) reported incident reports (none of which were from Canada) where patients with a Biotronik ICD experienced induced ventricular arrhythmia (VT and/or VF) following RF ablation of cardiac tissue using the Affera Sphere-9 catheter. Sphere-9 presents a higher current flow to the ICD or CRT-D which interacts with a safety feature built into Biotronik ICDs and CRT-Ds. The higher currents of the Sphere-9 catheter during RF application proceed to an unfavorable modulation of the coupled currents (approx. 40Hz frequency) that may be conducted via the implanted leads, and result in possible VT or VF induction.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime