Health Products

Spacelabs Healthcare Xhibit® Central Station

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Spacelabs Healthcare Xhibit® Central Station
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Spacelabs Healthcare Xhibit® Central StationNot applicable.96102

Issue

Spacelabs Healthcare has concluded that the use of the previously distributed splitter (part number 010-1975-00) can have an adverse effect on product performance, specifically around audible and visible alarms. Splitters may fail to perform as intended and result in degraded or complete loss of alarm audio and video. Additionally, customer's own use of unqualified splitters may pose the same risks.