Health Products

1788 4K Camera

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use, Health products - Medical devices - Gastroenterology and urology
Product: 1788 4K Camera
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
1788 4K Camera Head, C-Mount With Advanced Imaging Modality23H5197741788210105
1788 4K Camera Control Unit With Advanced Imaging ModalityNot applicable1788010000
1788 4K Camera Head With Integrated Coupler With Advanced Imaging ModalityMore than 10 numbers, contact manufacturer.1788610122

Issue

Stryker Endoscopy has received an increase in complaints for 1788 out of box failures due to bent connection pins. The bent pin may also damage the mating connector in the CCU. Stryker has investigated and identified a service repair issue which contributed to the bent pins noted with these devices.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime