| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| 1788 4K Camera Head, C-Mount With Advanced Imaging Modality | 23H519774 | 1788210105 |
| 1788 4K Camera Control Unit With Advanced Imaging Modality | Not applicable | 1788010000 |
| 1788 4K Camera Head With Integrated Coupler With Advanced Imaging Modality | More than 10 numbers, contact manufacturer. | 1788610122 |
Stryker Endoscopy has received an increase in complaints for 1788 out of box failures due to bent connection pins. The bent pin may also damage the mating connector in the CCU. Stryker has investigated and identified a service repair issue which contributed to the bent pins noted with these devices.