Health Products

Beckman Coulter AU Immunochemistry System - AU480 Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Beckman Coulter AU Immunochemistry System - AU480 Analyzer
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Beckman Coulter AU Immunochemistry System - AU480 Analyzer2021104332AU480
Beckman Coulter AU Immunochemistry System - AU480 Analyzer2021124493AU480
Beckman Coulter AU Immunochemistry System - AU480 Analyzer2022014627AU480
Beckman Coulter AU Immunochemistry System - AU480 Analyzer2021124487AU480
Beckman Coulter AU Immunochemistry System - AU480 Analyzer2022014611AU480

Issue

Beckman Coulter has determined that a software defect in AU480 chemistry analyzers using Microsoft® SQL server 2022 express may cause a DB connection error that renders alarmDB unreadable and prevents instrument startup. The issue is associated with transaction log auto-growth, and may cause delays in reporting patient test results to physicians.