Health ProductsType I recall

Endotracheal Tube, Cuffed

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Endotracheal Tube, Cuffed
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Endotracheal Tube, Cuffed21080114621-7333-25
Endotracheal Tube, Cuffed26030102741-7333-25

Issue

The 6FR suction catheter could not pass smoothly through the 2.5mm sizes of cuffed tracheal tube. This may lead to reintubations if the airway is unable to be suctioned.To date, there have been two complaints reported from united states customers where one required reintubation. No serious patient harm has occurred.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime