Health Products

SPECTRALIS® with Flex Module

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Ophthalmology
Product: SPECTRALIS® with Flex Module
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
SPECTRALIS® with Flex ModuleSpec-21986S2610
SPECTRALIS® with Flex ModuleSpec-24546S3610
SPECTRALIS® with Flex ModuleSpec-23776S3610
SPECTRALIS® with Flex ModuleSpec-13403S3610
SPECTRALIS® with Flex ModuleSpec-15400S3610

Issue

There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose. As a result, the Flex stand may tip over. Users shall schedule a service appointment as soon as possible so that the castor fastenings can be inspected and corrected.