· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| DxU 840m Iris | More than 10 numbers, contact manufacturer. | C76947 |
| iQ200 SPRINT 2008 Tested | More than 10 numbers, contact manufacturer. | 700-3325 |
| iQ200 SELECT 2008 Tested | More than 10 numbers, contact manufacturer. | 700-3345 |
| DxU 850m Iris | More than 10 numbers, contact manufacturer. | C49513 |
| iQ200 ELITE 2008 Tested | More than 10 numbers, contact manufacturer. | 700-3375 |
Beckman Coulter received customer complaints of lamina fluid leaks in iQ200 Series and DxU 840m/850m Analyzers. The investigation found that a structurally compromised rinse-line check valve may allow fluid to migrate internally and leak onto the table or floor, potentially resulting into exposure to physical or mechanical or chemical hazards.