Health Products

BIOFIRE® Joint Infection Panel – 30 Pack

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Pathology
Product: BIOFIRE® Joint Infection Panel – 30 Pack
Issue: Medical devices - Performance
What to do: Contact bioMérieux Canada Inc.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
BIOFIRE® Joint Infection Panel – 30 Pack0806526 (exp. 2027-03-24)RFIT-ASY-0138
BIOFIRE® Joint Infection Panel – 30 Pack0677326 (exp. 2027-03-13)RFIT-ASY-0138
BIOFIRE® Joint Infection Panel – 30 Pack1289226 (exp. 2027-05-05)RFIT-ASY-0138

Issue

Beginning April 5, 2026, elevated rates of KPC unexpected positives were observed in reagent final QC testing of the BIOFIRE® FILMARRAY® Pneumonia Panel, BIOFIRE® FILMARRAY® Pneumonia Panel Plus, and BIOFIRE® Joint Infection Panel, including in passing product that met final QC negativity requirements. Internal testing of failing product has reproduced KPC unexpected positive results, suggesting KPC nucleic acid (organism or synthetic) contamination in the pouch at levels that can lead to false positive results. Additional testing has confirmed that an isolated component, the outer primer pellet, is sufficient for producing unexpected positive KPC results.