| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Ingenia 1.5T | Not applicable. | 781315 |
| Ingenia 1.5T | Not applicable. | 1.5T |
| Ingenia 3.0T CX | Not applicable. | 781271 |
| Achieva 1.5T MRI System - Main Unit | Not applicable. | ACHIEVA 1.5T |
| Achieva 3.0T MRI System - Main Unit | Not applicable. | ACHIEVA 3.0T |
| Ingenia Ambition S | Not applicable. | 782108 |
| Ingenia Ambition S | Not applicable. | 782139 |
| Ingenia 3.0T | Not applicable. | 3.0T |
| Ingenia Ambition X | Not applicable. | 782109 |
| Ingenia Ambition X | Not applicable. | 781356 |
| Ingenia Ambition X | Not applicable. | 782138 |
| Ingenia Elition X | Not applicable. | 782107 |
| Ingenia Elition X | Not applicable. | 781358 |
| Intera 1.5T MRI System - Main Unit | Not applicable. | INTERA 1.5T |
Philips has become aware of a software issue affecting Philips Magnetic Resonance (MR) imaging systems operating with software versions R5 & R11, which may terminate an examination during the image reconstruction process. If the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory, the system will freeze, preventing image generation and requiring a system re-start.