· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Dxc 500i Clinical Analyzer | Software V. 1.3.5 or earlier | C63521 |
| Dxc 500i Clinical Analyzer | Software V. 1.3.5 or earlier | C63522 |
| Dxc 500 AU Clinical Chemistry Analyzer | Software V. 1.4.5 or earlier | C63520 |
| Dxc 500 AU Clinical Chemistry Analyzer | Software V. 1.4.5 or earlier | C63519 |
Beckman Coulter identified a software issue affecting DxC 500i and DxC 500 AU Analyzers when testing plasma samples. Following user configuration of a chemistry test to nondefault units, subsequent activation of a draft configuration in the system configuration page may revert units to the system default without notifying the user.