Health Products

Medline Single Sterile C-Section Drapes W/Incise

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Medline Single Sterile C-Section Drapes W/Incise
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Medline Single Sterile C-Section Drapes W/Incise31126040011FDYNJP6106
Medline Single Sterile C-Section Drapes W/Incise45025120094DYNJP6106
Medline Single Sterile C-Section Drapes W/Incise45025120253DYNJP6106
Medline Single Sterile C-Section Drapes W/Incise31126040011FDYNJP6108
Medline Single Sterile C-Section Drapes W/Incise45025120094DYNJP6108
Medline Single Sterile C-Section Drapes W/Incise45025120253DYNJP6108

Issue

Medline Industries, LP is issuing a correction notice for specific lot numbers of the Medline single sterile C-section drapes w/incise due to customer reports of C-section drapes being difficult to remove from patients, leaving adhesive residue on the patient's skin.