Health ProductsType I recall

Automated Impella® Controllers (AIC)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Automated Impella® Controllers (AIC)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Automated Impella® Controllers (AIC)IC135030042-0040-CA
Automated Impella® Controllers (AIC)IC132660042-0040-CA
Automated Impella® Controllers (AIC)IC111710042-0040-CA
Automated Impella® Controllers (AIC)IC111720042-0040-CA
Automated Impella® Controllers (AIC)IC17690042-0000-CA
Automated Impella® Controllers (AIC)IC58210042-0000-CA
Automated Impella® Controllers (AIC)IC58240042-0000-CA
Automated Impella® Controllers (AIC)IC94160042-0000-CA
Automated Impella® Controllers (AIC)IC94140042-0000-CA
Automated Impella® Controllers (AIC)IC11790042-0000-US

Issue

Abiomed has decided to initiate a voluntary medical device recall (removal) for design changes that were previously implemented through the Abiomed field service process. Abiomed performed a retrospective review of Field Change Notices (FCN) and determined that certain AIC consoles in the field did not receive the updates introduced by FCN 113.Failure to implement the applicable FCN 113 updates may result in the screen freezing during the Purge Cassette Change Wizard and the display of a false controller error.These defects could delay the initiation of support during device set-up and, in high-risk patient populations, may lead to temporary inadequate hemodynamic support or a temporary loss of hemodynamic support. Hospital inventory may continue to be used to ensure continuity of care. Not all units are affected by this issue.