Health Products

Philips Allura Xper Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Philips Allura Xper Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Allura Xper FD10Not applicable.ALLURA XPER FD10
Allura Xper FD10Not applicable.001443
Allura Xper FD10 (Product Of AlluraClarity Family)Not applicable.722026
Allura Xper FD20 (Product Of AlluraClarity Family)Not applicable.722028
Allura Xper FD20/10 (Product Of AlluraClarity Family)Not applicable.722029
Allura Xper FD20/15Not applicable.722058
Allura Xper FD20 OR Table (Product Of AlluraClarity Family)Not applicable.722035
Allura Xper FD20/20 (Product Of AlluraClarity Family)Not applicable.722038
Allura Xper FD10/10 (Product Of AlluraClarity Family)Not applicable.722027

Issue

Philips has identified a software issue that may cause the Flexvision software application to become unresponsive during clinical use. When this occurs, the FlexVision monitor may either:• turn blank and no longer display live images or other information shown on the FlexVision monitor; or• freeze on the last displayed screen and no longer update live images or other information shown on the flexvision monitor.A cold system restart is required to restore normal FlexVision display functionality.

Get told about the next health product recall.