Issue

Health Canada is informing Canadians and Canadian healthcare practitioners that Smiths Medical is conducting a North American voluntary recall of certain lot numbers (see below) of Bivona® Neonatal, Pediatric and Flextend Tracheostomy Tubes. A tracheostomy is an opening in the trachea made to help the patient breathe after surgery, or if there is a blockage in the airway. The kits are used to help keep the opening clear and clean.

According to the company, the recall is due to some customers experiencing difficulty disconnecting accessories from the connectors of the affected tubes. Some customers were also unable to disconnect the accessory or excessive force led to removal of the tube. If the tube is accidentally pulled-out, this can be a life-threatening situation. In some cases, emergency intervention by experts may be required to reinsert the tube to prevent suffocation.

Products affected

Smiths medical issues urgent device recall of certain bivona® neonatal, pediatric, and flextend tracheostomy tubes: Download PDF document

Who is affected

Canadians who have purchased or used Bivona® Neonatal, Pediatric and Flextend Tracheostomy Tubes.

What you should do

  • Consult your healthcare practitioner if you care for an infant or child that uses this device.
  • Identify all affected unused device in inventory and segregate it to a quarantine location, according to the company.
  • Contact Smiths Medical at 905-477-2000 ext. 135 with any questions regarding the Customer Information Bulletin.
  • View Airway TV Channel, according to the company.
  • Visit the Smiths Medical company's website, complete the "Confirmation Form" and return it to the company by Fax to 905-477-2144 or by email to melissa.lake@smiths-medical.com

Background

The tubes are class 2 medical devices and are used on the pediatric population in hospital as well as in the home setting. There has been twenty two complaints concerning the disconnection problem, but none have been reported in Canada.

What Health Canada is doing

Canadians can be confident that Health Canada is monitoring the voluntary recall by Smiths Medical. Should additional information be identified, Health Canada will take appropriate action and inform Canadians.

Report health or safety concerns

To report suspected adverse reaction to these or other health products, please contact the Canada Vigilance Program of Health Canada toll-free at 1-866-234-2345, or complete a Canada Vigilance Reporting Form and send to us using one of these methods:

Facsimile: 1-866-678-6789
Email: CanadaVigilance@hc-sc.gc.ca
Internet: MedEffect Canada
Mail: Canada Vigilance Program
Marketed Health Products Directorate
Ottawa, ON, Address Locator 0701C
K1A 0K9

Media enquiries

Health Canada
(613) 957-2983

Public enquiries

(613) 957-2991
1-866 225-0709
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime