This is duplicated text of a letter from Fisher & Paykel Healthcare.
Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
6 February, 2007
In support of our current retail-level recall, this public communication is being issued to remind Canadian end-users of the appropriate care and handling of their CPAP masks and connectors.
Fisher & Paykel Healthcare has become aware that a connector component (tab connector) used in a previous version of its CPAP masks may fail prematurely if it is not cleaned in accordance with our Instructions for Use.
We draw particular attention to the following Instructions for Use to be followed:
Before each use:
After each use:
Cautions:
Failure to adhere to these instructions may deteriorate or damage the mask. If your mask shows any signs of weakening or cracking, seek replacement immediately.
Instructions for use specific to your mask model may also be viewed at http://www.fphcare.com/osa/interfacesProducts.asp
The masks affected by this recall are Fisher & Paykel Healthcare Nasal, Oral, Direct Nasal and Full Face CPAP masks with tab connectors (Figure 1).
Masks manufactured since April 2006 have a new connector (Figure 2) instead of a tab connector. These newer masks are not affected by this recall.
Figure 1. Masks manufactured before April 2006 (previous version)

(a) Broken Tab Connector

(b) Tab Connector
Figure 2. Masks manufactured since April 2006 (new version)

New Connector
In the event of a tab breaking off, it may enter the CPAP system air path and there is an extremely remote possibility that the broken tab may then enter the patient's nasal cavity or mouth. Due to the low likelihood of a tab breaking off, the use of these CPAP masks with tab connectors is not likely to cause adverse health consequences.
A detailed list of affected models and lot numbers can be found in the Recall Fax Back Form.
Customers with the previous version connector as shown in figure 1(b) should continue using their mask. If your mask shows any signs of weakening or cracking, seek replacement from your homecare provider immediately.
For any inquiries related to this communication, please contact Fisher & Paykel Healthcare at:
United States
Fisher & Paykel Healthcare Inc.
15365 Barranca Parkway
Irvine, CA 92618
U.S.A.
Tel: 1 800 446 3908
info@fphcare.com
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
Health Canada
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and the Guidelines can be found on the Health Canada web site.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
E-mail: MHPD_DPSC@hc-sc.gc.ca
Tel: (613) 954-6522
Fax: (613) 952-7738
Sincerely,
original signed by
D. Michelle Garrett
Quality and Regulatory Manager
Fisher & Paykel Healthcare, Inc.
Fisher & Paykel Healthcare, Inc.
22982 Alcalde Drive
Suite 101
Laguna Hills
CA 92653
USA
Telephone: 1 800 446 3908
Facsimile: 1 949 470 3933
Website: www.fphcare.com
8 December, 2006
Voluntary Recall : Fisher & Paykel Healthcare CPAP Masks and Connectors
Fisher & Paykel Healthcare has become aware that a connector component used in a previous version of its CPAP masks may fail prematurely if improperly cleaned. Plastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with our Instructions for Use. In the unlikely event of a tab breaking off, it may enter the CPAP system air path.

Figure 1: Tab Connector

Figure 2: New (no tab) Connector
As a precautionary measure, Fisher & Paykel Healthcare has decided to conduct a retail level recall of CPAP masks and connectors with tabs (Figure 1) and replace them with our current no tab design (Figure 2). This will ensure that any new and replenishment product you provide will be CPAP masks or connectors using the new no tab design. See the attached fax back form for CPAP Mask models, spare part numbers and lot numbers that are affected by this voluntary recall.
Masks manufactured since April 2006 have a new connector instead of a tab connector. These masks are not subject to this voluntary recall.
In order to effectively support this voluntary recall please proceed as follows:

Figure 3. Lot No Indication
Complete the fax back form even if you have no stock which is subject to this voluntary recall, as we require this information to reconcile this process.
Please email recall@fphcare.com or call the number above if you have any questions relating to this letter, to obtain the Recall Fax Back Form, or to obtain a detailed list of affected models and lot numbers.
Fisher & Paykel Healthcare sincerely regrets any inconvenience caused.
Sincerely
Michelle Garrett
Quality and Regulatory Manager
Fisher & Paykel Healthcare Inc.