Archived – Important Medical Devices Safety Information – Product Correction and Removal on Your D-TRONplus Insulin Pump - For the Public

· Published by Health Canada · Official notice

Details

Alert type: Public Communication
Category: Medical Device
Issue: Medical Devices
Audience: General Public
Department: Health Canada
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Details

This is duplicated text of a letter from Disetronic Medical Systems, Inc.
Contact the company for a copy of any references, attachments or enclosures.

Notice about Health Canada advisories

Disetronic Medical Systems, Inc.
5151 Program Avenue
St. Paul, MN 55112-1014
800-280-7801
763-795-5200
Fax 763-795-5300
http://www.disetronic-usa.com

03-141
August 11, 2003
Canada

Issues

The U.S. Food and Drug Administration (FDA) inspected our insulin infusion pump manufacturing facility and determined that our manufacturing processes are not in substantial compliance with the regulations. In the absence of compliance with these regulations, the reliability of devices cannot be assured, which could potentially result in problems with the product leading to over and/or under infusion of insulin. Based on the evaluation of the observations with Health Canada, we have decided not to ship D-TRONplus insulin pumps to any new customers until we have improved our manufacturing processes to adequately ensure the reliability of our pumps for their intended use. Some of the problems that have been reported include stuck piston rods and punctured buttons. Pumps that do not perform reliably as expected may lead to hypoglycemic or hyperglycemic events, the symptoms and actions for these possible events are described on page three of this letter.

Health Canada has asked that we inform you of the problems with this pump. You should discuss this information with your healthcare professional so that you can make an informed, joint decision regarding the continued use of your D-TRONplus pump. After discussing this with your healthcare professional, if you wish to discontinue use of your pump, have your pump replaced, or have questions regarding your pump, please refer to the "Actions" section of this letter.

Important Information

Please review the following information carefully.

Stuck Piston Rod

  • Warning: A stuck piston rod can cause under delivery of insulin, resulting in hyperglycemia. Anytime you have an unexplained high blood sugar, or with any reading over 16.65 mmol/L on an insulin pump, take a second reading to confirm the result. If the result is still high, suspect an insulin delivery problem. If there is a clear reason for the high reading, follow the advice of your healthcare professional and recheck your sugar in one hour to ensure it is being corrected.
  • Note: If the piston rod sticks, an "Alert A4: Cartridge/Adapter Alert" or an "Error E6: Mechanical Error" message may be given. If this occurs, you could get a high glucose reading.

If there is no clear reason for the high glucose reading, or the second blood sugar remains high:

  1. Check for pump or site-related problems. This can be done by following the procedure in the D-TRONplus Reference Manual, 4.5 Prime the infusion set.
  2. Check your blood or urine for ketones. Ketones suggest you have a problem with insulin delivery from your pump or a serious illness. If you have ketones present, contact your healthcare professional. If you suspect a problem with your pump, please call 1-800-688-4578.

Punctured Key(s)

A punctured key is one with a crack or hole in the key pad. Punctured keys can result from using a sharp item or your fingernails when pressing the key pad.

If you, along with your healthcare professional, decide to continue using your D-TRONplus pump, we would like to make you aware that punctured keys can allow water to enter your pump. You must not use a pump with punctured keys in or around water.

  • Caution: We would like to remind you to never press any key on your pump with a sharp item or with your fingernails. Only use the pad of your finger to prevent puncturing the keys. Punctured keys can allow water to enter the pump.
  • Caution: If your D-TRONplus pump has punctured keys, avoid any contact with water as your pump may or may not alarm if water has entered the pump. If you obtain an "Error E7: Electronic Error" and you detect water in your pump or it has been exposed to water, DO NOT follow the instructions in the D-TRONplus pump Reference Manual. Instead, call the number on the back of your pump to arrange for a replacement.

For your convenience, we are including pages copied from the D-TRONplus Reference Manual describing the alarms mentioned above and the corrective actions.

Precautions

If your pump has punctured keys avoid contact with water and contact us to arrange for a replacement. If your pump has punctured keys and has been in contact with water or if you detect water in the pump casing, discontinue use of your pump and take the following steps:

  • Measure your blood glucose value immediately and correct it, if necessary, using an insulin injection or your backup insulin pump.
  • If you are unable to correct your blood glucose value, contact your healthcare professional immediately or get transportation to the nearest emergency room.
  • If you are able to correct your blood glucose value, continue therapy with your backup insulin delivery device and follow the instructions for returning your pump. If you are unsure if your blood glucose is correct, call your healthcare professional.
  • Call the toll-free number (1-800-688-4578) on the back of your pump for troubleshooting help or to make arrangements to return your D-TRONplus pump for inspection or replacement.

Backup System

You should always have insulin and syringes on hand as a backup to your insulin pump. If you do not have these supplies, please contact your healthcare professional for a prescription and obtain these supplies. If you purchased a two-pump system and are still awaiting a second pump, we will not be able to ship your second pump until we have completed certain corrective actions in our manufacturing process. When we have pumps available to ship, we will contact you to arrange shipment.

Symptoms of Hypoglycemia and Hyperglycemia

Symptoms of hypoglycemia include:

  • Shakiness
  • Dizziness
  • Sweating
  • Hunger
  • Headache
  • Pale skin color
  • Seizure
  • Difficulty paying attention
  • Confusion
  • Clumsy or jerky movements
  • Tingling around mouth
  • Sudden moodiness or behavior changes, such as crying for no apparent reason

Symptoms of hyperglycemia include:

  • Fatigue/drowsiness
  • High levels of sugar in the urine
  • Increased thirst
  • Frequent urination

Actions

  • You should always have insulin and syringes on hand as a backup to your insulin pump.
  • If you are experiencing any of the symptoms described above, please contact your healthcare professional.
  • Review this information carefully.
  • Inspect your pump for punctured keys and the presence of water.
  • If water is present refer to the Precautions section of this communication for the appropriate course of action.
  • If you find punctured keys, call the number (1-800-688-4578) on the back of your pump to arrange for a replacement.
  • We will continue to provide you with any pump service you may require, as well as all of your infusion set and other disposable needs.
  • If you experience any problem with your pump, promptly inform us by calling 1-800-688-4578, available 24 hours a day.
  • Continue your daily regimen of blood glucose monitoring and insulin dosage per your healthcare professional's orders.
  • After discussing this information with your healthcare professional, if you wish to discontinue use of your pump, have your pump replaced, or have questions or concerns regarding the continued use of your pump or about this letter, please call 1-866-307-2344.
  • We have also included answers to frequently asked questions on our website (http://www.disetronic-usa.com).

Questions

We deeply regret any inconvenience this situation may cause. This notification is being made with the knowledge of Health Canada. Please contact us at 1-866-307-2344 if you have questions about the information contained in this communication.

Any suspected adverse incident can also be reported to:
Health Products & Food Branch Inspectorate
HEALTH CANADA
Address Locator: 3002C
OTTAWA, Ontario, K1A 0K9
Tel: The Medical Devices Hotline 1-800-267-9675