This is duplicated text of a letter from Disetronic Medical Systems, Inc.
Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
Disetronic Medical Systems, Inc.
5151 Program Avenue
St. Paul, MN 55112-1014
800-280-7801
763-795-5200
Fax 763-795-5300
http://www.disetronic-usa.com
03-141
August 11, 2003
Canada
Issues
The U.S. Food and Drug Administration (FDA) inspected our insulin infusion pump manufacturing facility and determined that our manufacturing processes are not in substantial compliance with the regulations. In the absence of compliance with these regulations, the reliability of devices cannot be assured, which could potentially result in problems with the product leading to over and/or under infusion of insulin. Based on the evaluation of the observations with Health Canada, we have decided not to ship D-TRONplus insulin pumps to any new customers until we have improved our manufacturing processes to adequately ensure the reliability of our pumps for their intended use. Some of the problems that have been reported include stuck piston rods and punctured buttons. Pumps that do not perform reliably as expected may lead to hypoglycemic or hyperglycemic events, the symptoms and actions for these possible events are described on page three of this letter.
Health Canada has asked that we inform you of the problems with this pump. You should discuss this information with your healthcare professional so that you can make an informed, joint decision regarding the continued use of your D-TRONplus pump. After discussing this with your healthcare professional, if you wish to discontinue use of your pump, have your pump replaced, or have questions regarding your pump, please refer to the "Actions" section of this letter.
Important Information
Please review the following information carefully.
Stuck Piston Rod
If there is no clear reason for the high glucose reading, or the second blood sugar remains high:
Punctured Key(s)
A punctured key is one with a crack or hole in the key pad. Punctured keys can result from using a sharp item or your fingernails when pressing the key pad.
If you, along with your healthcare professional, decide to continue using your D-TRONplus pump, we would like to make you aware that punctured keys can allow water to enter your pump. You must not use a pump with punctured keys in or around water.
For your convenience, we are including pages copied from the D-TRONplus Reference Manual describing the alarms mentioned above and the corrective actions.
Precautions
If your pump has punctured keys avoid contact with water and contact us to arrange for a replacement. If your pump has punctured keys and has been in contact with water or if you detect water in the pump casing, discontinue use of your pump and take the following steps:
Backup System
You should always have insulin and syringes on hand as a backup to your insulin pump. If you do not have these supplies, please contact your healthcare professional for a prescription and obtain these supplies. If you purchased a two-pump system and are still awaiting a second pump, we will not be able to ship your second pump until we have completed certain corrective actions in our manufacturing process. When we have pumps available to ship, we will contact you to arrange shipment.
Symptoms of Hypoglycemia and Hyperglycemia
Symptoms of hypoglycemia include:
Symptoms of hyperglycemia include:
Actions
Questions
We deeply regret any inconvenience this situation may cause. This notification is being made with the knowledge of Health Canada. Please contact us at 1-866-307-2344 if you have questions about the information contained in this communication.
Any suspected adverse incident can also be reported to:
Health Products & Food Branch Inspectorate
HEALTH CANADA
Address Locator: 3002C
OTTAWA, Ontario, K1A 0K9
Tel: The Medical Devices Hotline 1-800-267-9675