Health ProductsPublic Communication

Archived - For the Public - Urgent: Medical Device Recall - Models Affected: S8 Compact, S8 Escape, S8 Elite and S8 AutoSet Vantage - ResMed Corp

· Published by Health Canada · Official notice

Details

Alert type: Public Communication
Category: Medical Device
Audience: General Public
Department: Health Canada
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Details

This is duplicated text of a letter from ResMed Corp.
Contact the company for a copy of any references, attachments or enclosures.

Notice about Health Canada advisories

Public Communication
Health Canada Endorsed Important Safety Information on certain S8 flow generator devices

May 17, 2007

Urgent: Medical Device Recall

Models Affected: S8 Compact, S8 Escape, S8 Elite and S8 AutoSet Vantage

Dear Patient:

I am writing to inform you about a recall initiated by ResMed on certain S8 flow generator devices. The power supply connection in certain S8 devices manufactured before May 15, 2006 may be susceptible to fatigue and failure. ResMed has implemented a program to replace the affected S8 devices. According to our records you have one of the affected S8 devices.

As background, in the majority of the cases of reported failure, the device has simply stopped operation. There have been a limited number of instances where the short circuit caused thermal damage to the device. To date, there have not been any reported instances of significant property damage or patient injury. ResMed has determined that the risk of property damage or patient injury from electrical failure is remote.

You may continue to use your S8 flow generator in advance of receipt of a replacement device. To begin the replacement process, we ask that you:

  1. Contact us at 888-899-8991 to verify that the serial number on your device is part of the replacement program.
    You can locate the 11-digit serial number on the bottom of your device. Before turning your device over, please disconnect your H3i humidifier, if you have one.
  2. Once we have verified that your device is included in the replacement program we will schedule an appointment for someone to come to your home with the new S8 flow generator. While they are at your home they will transfer all required prescription information from the old device to the new S8 device. We anticipate that this process will take approximately 30 minutes. Your old S8 flow generator will also be collected at that time.

As with any electrical device, please continue to ensure that your S8 device is placed on a hard clean surface and that the area around the device is clear. The following precautions may further reduce risk in advance of receiving your replacement device:

  • Do not use supplemental oxygen with an affected device, until a replacement S8 device has been provided. If you currently use supplemental oxygen, please immediately call 888-899-8991 for a replacement S8 device.
  • Discontinue use of the device if there are any signs of electrical failure such as: intermittent power, cracking sounds, sparking or charred smell. If this does occur, please immediately call 888-899-8991 to return the unit and receive a replacement S8 device.

If you have any questions regarding this recall program, please feel free to contact us at 888-899-8991.

ResMed Corp
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153
Australia
Tel: 1-888-899-8991

Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
Health Canada
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675

The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.

For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
E-mail: mhpd_dpsc@hc-sc.gc.ca
Tel: 613-954-6522
Fax: 613-952-7738

Sincerely,

original signed by

Jon Brockman
Director, Quality Assurance & Regulatory Affairs
ResMed Corp

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime