Subject: Mallinckrodt, a division of Nellcor/Tyco Healthcare Issues Recall of TracheoSoft XLT Extended Length Tracheostomy Tube
Pleasanton, Calif. - August 16, 2004 - On July 8, 2004, Nellcor, part of Mallinckrodt, a Tyco Healthcare company, initiated a voluntary recall of the Shiley® TracheoSoft® XLT Extended Length Tracheostomy Tube. This product can be used in the hospital or home-care environment to provide tracheal access in patients for airway management. The recall was prompted by reports of the outer cannula separating from the swivel neck plate. While the number of customer reports of cannula separation is very low, if this failure mode does occur, the outer cannula could travel into the patient's airway.
Accordingly, in the interest of patient safety, the recall extends to all lots of the following products:
| Model Number | Description |
|---|---|
| 72110-050 | Size 5, Proximal Extension, Uncuffed |
| 72110-060 | Size 6, Proximal Extension, Uncuffed |
| 72110-070 | Size 7, Proximal Extension, Uncuffed |
| 72110-080 | Size 8, Proximal Extension, Uncuffed |
| 72120-050 | Size 5, Proximal Extension, Cuffed |
| 72120-060 | Size 6, Proximal Extension, Cuffed |
| 72120-070 | Size 7, Proximal Extension, Cuffed |
| 72120-080 | Size 8, Proximal Extension, Cuffed |
| 73110-050 | Size 5, Distal Extension, Uncuffed |
| 73110-060 | Size 6, Distal Extension, Uncuffed |
| 73110-070 | Size 7, Distal Extension, Uncuffed |
| 73110-080 | Size 8, Distal Extension, Uncuffed |
| 73120-050 | Size 5, Distal Extension, Cuffed |
| 73120-060 | Size 6, Distal Extension, Cuffed |
| 73120-070 | Size 7, Distal Extension, Cuffed |
| 73120-080 | Size 8, Distal Extension, Cuffed |
| 77100-050 | Size 5, XLT, Disposable Inner Cannula |
| 77100-060 | Size 6, XLT, Disposable Inner Cannula |
| 77100-070 | Size 7, XLT, Disposable Inner Cannula |
| 77100-080 | Size 8, XLT, Disposable Inner Cannula |
The company received two reports in which the cannula separation was purportedly associated with a patient death. Neither of these reports originated in Canada. Patients or their families should contact their doctor to determine if they are using the recalled product. Those TracheoSoft XLT Tracheostomy Tubes currently in use should be exchanged, when practical, with an alternative model at the discretion and judgment of the patient's physician.
Institutions with affected product have received packages outlining the recall process. The company is contacting those locations to insure compliance with the recall instructions. Health Canada and officials in other affected countries have been apprised of this recall action.
This action does not affect any other model of Shiley tracheostomy products. "Our foremost concern is patient safety, which is why we voluntarily recalled the product," said Michael Frazzette, president of Tyco Healthcare Canada. "We regret any disruption caused for physicians and patients, and we are investigating the cause of the malfunction to prevent recurrence."
Clinicians and patients with inquiries should contact Tyco Healthcare's Customer Service Department at 1-877-664-8926, option 1.
Tyco Healthcare, a business segment of Tyco International Ltd., is a leading manufacturer, distributor and servicer of medical devices worldwide. Its broad portfolio includes disposable medical supplies, monitoring equipment, medical instruments and bulk analgesic pharmaceuticals, sold under such names as Auto Suture, Kendall, Mallinckrodt, Nellcor, Puritan Bennett, United States Surgical, Valleylab and others.
The identification, characterization, and management of medical device-related adverse reactions are dependent on the active participation of health care professionals in adverse reaction reporting programmes. Any occurrences of serious and/or unexpected adverse reactions should be reported to Tyco Healthcare or Health Canada at the following addresses:
Tyco Healthcare Canada
7500 Trans Canada Highway,
Pointe-Claire, Québec
H9R 5H8
1-877-664-8926, option 1
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
Health Canada
Address Locator: 3002C
Ottawa, Ontario K1A 0K9
Tel: The Medical Devices Hotline 1-800-267-9675
For other inquiries & attachments: please refer to contact information.
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
original signed by
Marie-Claude Boudrias
Director QA/RA