· Published by Health Canada · Official notice
This is duplicated text of a letter from Disetronic Medical Systems, Inc.
Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
March 31, 2006
Subject: URGENT PRODUCT RECALL ACCU-CHEK™ Ultraflex infusion sets: Potential for full or partial separation at the luer lock-tubing connection
This recall applies to all ACCU-CHEK Ultraflex infusion sets.
In the course of normal daily use, ACCU-CHEK Ultraflex Infusion Sets can bend at the luer lock-tubing connection. This bending can lead to full or partial separation at the luer lock-tubing connection. In the event that a full or partial separation occurs, it is possible that insulin could leak from the infusion set tubing causing an interruption of insulin delivery, which can and has contributed to hyperglycemia.
We are taking this action because we have recently received an increase in complaints regarding fully or partially separated luer lock-tubing connections with ACCU-CHEK Ultraflex infusion sets. We are conducting an ongoing investigation into these issues and will update you as new information becomes available.
YOU MUST check your infusion set at the luer lock-tubing connection during use at least every 3 hours and before bedtime. A full or partial separation may result in leakage, causing interruption of insulin delivery and elevated blood glucose levels.
If you notice a full or partial separation, change your infusion set immediately and check your blood glucose to ensure that your blood glucose level is within an acceptable range as defined by your healthcare team.

Illustrations above depict the ACCU-CHEK Ultraflex luer lock-tubing connection.
An interruption of insulin delivery for this or any other reason can lead to hyperglycemia. The symptoms of hyperglycemia are:
If you experience these symptoms please check your blood glucose to ensure that your blood glucose level is within an acceptable range as defined by your healthcare team and follow the medical advice given to you by your Healthcare Professional or contact your physician.
This Urgent Product Recall is being made with the knowledge of Health Canada. If you have experienced a problem with any of these infusion sets, please call 1-800-267-9675. If you have further questions about the information contained in this Urgent Product Recall, please contact Pump Support at 1-800-688-4578, 24 hours a day, 365 days a year.
Disetronic Medical Systems, Inc.
11800 Exit 5 Parkway
Suite 120
US - Fishers, IN 46038
Tel: +1 317 570 5100
Toll-free: +1 800 280 7801
Fax: +1 317 570 5300
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
Email: MHPD_DPSC
Tel: (613) 954-6522
Fax: (613) 952-7738