· Published by Health Canada · Official notice
Notice about Health Canada advisories
April 21, 2004
To: Hospital Chief of Medical Staff
Hospital Infection Control Practitioners / Department
Provincial and Territorial Ministries of Health
Please distribute to the relevant Departments of Surgery and other involved professional staff and Post this NOTICE in your institution.
Subject: Inadequate Cleaning/Sterilization of EZ Clean Monobloc ACETABULAR REAMER a reuseable medical device
Health Canada has been made aware of cases where a reusable medical device may not be adequately cleaned and sterilized prior to use on patients. Health Canada has conducted a risk assessment and, based on the present situation, has concluded that the risk of transmission of blood borne disease using an inadequate cleaning and sterilization procedure is low. Although the risk is considered to be low, Health Canada recommends that hospitals take appropriate measures to reduce this risk to patient safety.
The device is being sold in Canada under the tradename EZ Clean Monobloc Acetabular Reamer Shaft. This shaft can also be included in the EZ Clean Reamer System. It is manufactured by Precimed S.A, rue de l'Echelette 7, CH-2534, Orvin, Switzerland and imported into Canada by Zimmer of Canada Limited, 2323 Argentia Road, Mississauga, Ontario, under the product number 1206-90-10.
The Medical Devices Regulations require the manufacturer to obtain a licence to sell this medical device in Canada. Please be advised that the manufacturer has not obtained a licence. Should you become aware that your facility is in possession of this unlicenced device we ask that you provide, in writing, a letter indicating any problems which have been encountered in using the device. This information should be sent to the address indicated at the end of this notice.
Information has also been presented to Health Canada that complete instructions for the proper cleaning and sterilization of the device prior to reuse may not have been provided at the time of purchase. Health Canada will continue to review this matter with users and suppliers and will take appropriate actions to address this issue should this be the case.
Health Canada would like to take this opportunity to recommend to hospitals who clean, disinfect and sterilize reusable medical devices that they should have in place:
Any situation where the cleaning, disinfecting and sterilizing instructions for a device are not included with the device should be reported to Health Canada at the following address:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 3002C
Ottawa, Ontario K1A 0K9
Tel: The Medical Devices Hotline 1-800-267-9675
For other inquiries: please refer to contact information.
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.