· Published by Health Canada · Official notice
Notice about Health Canada advisories
April 16, 2004
To: Hospital Chief of Medical Staff
Please distribute to the relevant Departments of Nursing, Surgery, Emergency Medecine, Pharmacy, Pediatrics, Anesthesia, Geriatrics, Internal Medicine, Dentistry, Intensive Care and other involved professional staff of nursing homes and nursing associations. Post this NOTICE in your institution.
Subject: Health risks associated with use of INFUSION PUMPS
Health Canada is aware of numerous injuries and some deaths associated with the use of infusion pumps, including general purpose pumps, enteral pumps, insulin and patient-controlled analgesia (PCA) pumps. The purpose of this notice is to provide information about the nature of these incidents and recommendations for reducing their frequency.
Between 1987 and March 2003, Health Canada received reports of 425 separate incidents involving infusion pumps. A few of these incidents did not occur in Canada. Of the 425 incidents, 23 resulted in death, 135 resulted in injury, and 127 could potentially have led to death or injury. Of these, 20 deaths and all 135 injuries were caused or suspected to have been caused by the pump. These include incidents that were caused by user error, but not those that may have been caused by deliberate misuse. Further statistics are given in the charts below.
Recently, concern over such adverse incidents has led several organizations to publish articles on the topic. In May 2001, ECRI published a review of PCA pumps1Footnote 1 with evaluation criteria that can probably apply to other types of pumps and should be considered before purchasing and using a pump.
In July 2003, the Institute for Safe Medication Practices Canada (ISMP Canada), issued a Safety BulletinFootnote 2 which highlighted the lack of free-flow protection mechanisms, which can result in serious consequences if the user fails to use the manual roller clamp on the infusion line. Two other bulletins from ISMPFootnote 3 provide information about safety issues relating to patient-controlled analgesia (PCA) pumps.
A national hospital surveyFootnote 4, conducted in 2003 by ISMP Canada and co-ordinated through the Canadian Healthcare Association, attempted to identify problems users have experienced with pumps in Canada. This survey indicated that a number of facilities still use pumps without adequate free flow protection. Other problems being cited include incorrect flow rate, dosage errors and air sensor problems. The detailed results of the survey will soon be published on the ISMP web site.
Health Canada's review of incident data has also revealed reports of misloading of the IV tubing in the pump. In some cases, the pump's misloading sensors may fail to identify this situation, leading to over or under infusions without alarm.
In March, 2003, the US Joint Commission on Accreditation of Healthcare Organizations (JCAHO) indicated, as part of its National Patient Safety Goals, that it would assess accredited facilities to ensure that free-flow protection is present on all general-use and PCA intravenous infusion pumpsFootnote 5.
In light of the reported incidents and above articles, Health Canada makes the following recommendations:


Health Canada asks that you share these recommendations with your staff or membership and encourage their implementation in the interest of patient safety.
For additional information on the technical content of the Notice, please contact the Manager, Device Surveillance Division, Medical Devices Bureau at 613-954-0287. If you have any concerns, problems or complaints pertaining to INFUSION PUMPS, you may report them to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 3002C
Ottawa, Ontario K1A 0K9
Tel.: The Medical Devices Hotline 1-800-267-9675
For other inquiries: please refer to contact information.
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
ECRI Health Devices 30 (5), May 2001
ISMP Canada Safety Bulletin, July 2003, volume 3, issue 7, "Infusion Pumps - Opportunities for Improvement", ISMP Canada and HIROC
ISMP Medication Safety Alerts, volume 8, issue 14, July 10, 2003, and volume 8, issue 15, July 24, 2003
http://www.ismp-canada.org/ipss.htm
JCAHO's 2004 National Patient Safety Goal to improve the safety of infusion pump use by ensuring free-flow protection on all general-use patient-controlled analgesia pumps, http://www.jcaho.org/accredited+organizations/
patient+safety/04+npsg/04_npsg.htm
ECRI Special Report SR0018, March 21, 2003.