Dear Client,
Recently, BHM Medical received an incident report concerning the Ergolift actuator. The incident involved a movement of the retainer ring due to wearing at the end of the stroke mechanism. If the wearing is not detected early, there is a possibility that the actuator may break causing a free fall.
The report showed that the Ergolift actuators are fully secure when used within their maximum extension length. However, on a few of the actuators, we found that some wearing of the retainer ring may have occurred when the actuator was frequently extended to the end of the stroke. Frequent over extension may weaken the actuator increasing the risk of actuator breakage during a patient transfer.
The equipment in question is the actuator LA-32 (black motor). These actuators were installed on Ergolifts and carry serial numbers ERLI-899 and lower. Our database indicates that you have purchased one of these Ergolifts.
We suggest one of the following options to ensure the reliability and the safety of your lift.
To inform BHM of the option you choose, please contact BHM Customer Service at 1-800-868-0441.
For information regarding the cost of each of the above options, or for any other questions or concerns, please feel free to communicate with our Customer Service Department, at 800-868-0441.
Sincerely,
Customer Service Department
BHM Medical Inc.