Health ProductsNotice to Hospitals

Risk of Inadvertent Disconnection of Hemodialysis Access Devices – Notice to Hospitals

· Published by Health Canada · Official notice

Details

Alert type: Notice to Hospitals
Category: Medical Device
Issue: Important Safety Information
Audience: Hospitals
Department: Health Canada
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Details

Notice about Health Canada advisories

 
[Text of letter begins]

Notice to Hospitals - Health Canada Endorsed Important Safety Information on Hemodialysis Access Devices

January 19, 2005

To: Hospital Chief of Medical Staff

Please distribute to the Hemodialysis Units and other involved professional staff and post this Notice in your institution.

Subject: Risk of inadvertent disconnection of hemodialysis vascular access devices

Health Canada is concerned about reports of accidental disconnection of vascular access devices during hemodialysis treatment. This is a rare but potentially serious problem leading to rapid, significant blood loss and its sequelae.

Vascular access devices include large bore needles and central venous catheters as well as any other device used to connect the bloodlines of the dialysis machine to the patient's vascular system.

Hemodialysis machines are equipped with venous pressure monitors that alarm when the value of the pressure in the venous bloodline falls below the limit set by the user. However, the venous pressure drop resulting from the inadvertent disconnection of a vascular access device may not be large enough to trigger an alarm. For example, a disconnection of the fistula needle may not trigger the alarm because the major source of resistance in this area of the circuit is the needle itself. Therefore, venous pressure monitors should be viewed only as one of several safeguards used to detect accidental disconnection.

Health Canada recommends the following:

  • review your centre's specific policies and procedures for ensuring the security of the access devices and bloodlines
  • verify that the bloodline is firmly attached to the access device and properly secured
  • ascertain that all connections in the blood circuit are properly closed by staying with the patient during the initial 3-5 minutes of dialysis treatment
  • monitor the patient carefully and regularly
  • ensure that the vascular access remains fully visible (not covered by clothing or other items)
  • set a reasonable venous pressure alarm limit and do not disable the alarm

Health Canada depends on health care professionals to report adverse events related to medical devices. Problems should be reported to Health Canada at the following address:

Any suspected adverse incident can be reported to:
Health Products and Food Branch Inspectorate
Health Canada
Address Locator: 3002C
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675

For other inquiries: please refer to contact information
Marketed Health Products Directorate
email: MHPD_DPSC@hc-sc.gc.ca
Tel: (613) 954-6522
Fax: (613) 952-7738

The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.

[Text of letter ends]
 

References:

  1. Baxter Important Clinical Practice Reminder, October 6, 2004.
  2. Ottawa Hospital- Nephrology Program- Nephrology Practice Alert #36, September 16, 2004.