April 26, 2005
To: Hospital Chief of Medical Staff
Please distribute to the relevant Departments of Radiology and Pharmacy and other involved professional staff and post this NOTICE in your institution.
Subject: Association of transdermal drug patches with thermal burns during magnetic resonance imaging procedures
Health Canada is aware of international reports of patient injury associated with the use of transdermal medication patches during magnetic resonance imaging (MRI) procedures. Although no similar incidents have been reported in Canada, this notice is being issued to increase awareness of this problem and provide relevant recommendations.
Transdermal drug patches are adhesive patches used to deliver a dose of drug through the skin at a controlled rate over a period of time. Some transdermal patches contain an aluminum or foil backing in the layer farthest from the skin. This backing helps to contain the drug in the patch's reservoir while it is slowly administered transdermally.Footnote 1
MRI systems produce radio wave pulses of energy. If conductive materials are placed in the MRI field, a concentration of electrical currents sufficient to cause excessive tissue heating could be induced. Though the metallic components found in some transdermal patches are not ferromagnetic, they are conductive and this led to patient burns during MRI procedures in the incidents reported internationally.Footnote 2
In light of this information, Health Canada recommends the following:
Health Canada depends on health care professionals to report adverse incidents related to medical devices. Any serious and/or unexpected adverse incident related to medical devices should be reported to Health Canada at the following address:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 3002C
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675
For other inquiries, please refer to contact information:
Marketed Health Products Directorate (MHPD)
MHPD_DPSC@hc-sc.gc.ca
Tel/Tél: (613) 954-6522
Fax/Télé: (613) 952-7738
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
Karch AM. Don't Get Burnt by the MRI. Am J Nurs 2004; 104(8): 31.
ISMP Medication Safety Alert. April 8, 2004.