Notice about Health Canada advisories
July 20, 2005
To: Hospital Chief of Medical Staff
Please distribute to the relevant Departments of Surgery, Emergency Medecine, Pediatrics, Anesthesia, Geriatrics, Internal Medicine, Nursing, Intensive Care and other involved professional staff and post this NOTICE in your institution.
Subject: Potential patient entrapment in Vail Enclosed Beds
Health Canada has recently become aware of safety concerns with all models (500, 1000 and 2000) of Vail enclosed beds sold by Vail Products based in Toledo, Ohio. There have been some entrapment incidents and deaths reported in the United States, but none in Canada. Due to actions taken by the US Food and Drug Administration (USFDA), Vail Products has decided to cease operations. Information regarding the USFDA’s action is described the Food and Drug Administration web site.
The actions by the USFDA affects all models of Vail enclosed beds, i.e. models 500, 1000 and 2000. There are reports of 30 entrapment incidents, including 8 deaths, with more than half involving children. Adverse events are often under-reported.
Health Canada has performed a safety evaluation of this bed and has concluded that provided you have a retrofit kit (an accessory attached to all sides, head and foot of the bed to improve safety) properly installed on the bed and follow recently released revised user instructions, the bed does not present an unreasonable risk. Any bed manufactured after July 25, 2002, would have had this improvement provided at time of manufacture. For other beds, the company made the kit available to users. Vail’s records indicate that less than half the beds sold in Canada have had the kit installed. Revised labelling and instructions for use have been provided by Vail to all Canadian and US customers on June 17, 2005. Users who opt to continue using the beds are advised to follow these new instructions.
Health Canada advises customers that since they can no longer obtain the retrofit kits or revised user instructions from Vail, they should follow Health Canada’s recommendations in this document as well as the revised instructions for use (see item 11 under Safety Precautions below).
If your bed is not equipped with the manufacturer’s retrofit kit, we advise you to use another bed system. Should you decide to continue using the bed without a retrofit kit installed, we urge you to pay particular attention to the safety precautions further below and seek medical advice on the continued use of the bed. These safety precautions should be followed whether you have the kit installed or not.
To determine if you have a retrofit kit, please refer to Figures 1 to 3 in the image gallery at the end of this letter. Note: model 500 beds were provided at time of manufacture with only the end rails. Shim bars for all model beds can be obtained by contacting Mr. Rick Vail at R.C. Medical at (419) 862-2883. Please note that even with the kit installed, this bed, like many medical beds, can present some risk of entrapment. There have been some reports in the U.S. of incidents, including deaths, with beds equipped with the kits, although many were associated with improper kit installation or improper use.
Health Canada recommends that you ensure that the retrofit kit, if shipped separately, was properly installed. There have been reports of improper assembly of the bed itself. Please verify that all assembly instructions were properly followed. R.C. Medical may provide some assistance with questions about installation.
Health Canada continues to monitor the situation and will provide further communication if appropriate. Should you have any incidents to report with this bed, please contact the Health Products and Food Branch Inspectorate at 1-800-267-9675.
Should you have any questions or concerns, you may contact Dr. Roland Rotter, Director of the Medical Device Bureau, Health Canada at telephone 613-957-4786.
The identification, characterization, and management of medical device-related adverse incidents are dependent on the active participation of health care professionals in adverse incident reporting programmes. Any adverse incident should be reported to Health Canada at the following address:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 3002C
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675
For other inquiries, please refer to contact information:
Medical Device Bureau Enquiries
E-mail: mdb_enquiries@hc-sc.gc.ca
Tel: (613) 957-4786
Fax: (613) 957-7318
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.