Health ProductsDear Healthcare Professional Letter

Revised safety information on Guidant Implantable Cardiac Defibrillators (VENTAK PRIZM AVT, VITALITY AVT, and CONTAK RENEWAL AVT) – Guidant Canada Corporation – For Health Professionals

· Published by Health Canada · Official notice

Details

Alert type: Dear Healthcare Professional Letter
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals
Department: Health Canada
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Details

This is duplicated text of a letter from Guidant Canada Corporation
Contact the company for a copy of any references, attachments or enclosures.

Notice about Health Canada advisories

 
[Text of letter begins]

Health Canada Endorsed Revised Safety Information on Guidant Implantable Cardiac Defibrillators (VENTAK PRIZM AVT, VITALITY AVT and CONTAK RENEWAL AVT)

July 22, 2005

Subject: Update Regarding VENTAK PRIZM AVT®, VITALITY® AVT and CONTAK RENEWAL® AVT

Dear Health Care Professional,

This letter updates previous safety information regarding all serial numbers of VENTAK PRIZM AVT, VITALITY AVT, and CONTAK RENEWAL AVT. Our records indicate that you have implanted or are monitoring patients that have these devices. Guidant is revising its original recommendations set forth in the June 20, 2005 letter because new information indicates that one of the original recommendations can increase the risk of a latching event.

As described in a June 20, 2005 communication to physicians, Guidant has determined that the atrial therapy (AVT) subgroups of certain Guidant ICD and CRT-D product families are subject to a condition in which functional “latching” limits available therapy. At that time, two occurrences had been confirmed out of approximately 20,950 devices implanted to date.

On July 11, 2005, a third AVT latching event was reported in the United States. Guidant immediately began analysis and has determined that this event occurred despite Atrial Episode Data Storage being programmed to 0%. This resulted in a latched state of continuous pacing at approximately 120 pulses per minute. This third event, similar to the first two events, resulted in no apparent patient injury beyond device replacement. Additional events, including a possible injury, are being evaluated.

Clinical Implications
Latching of AVT devices will suspend detection and treatment of atrial and ventricular arrhythmias. Telemetry, programming and magnet response are not available. Brady pacing may continue, but will not be programmable and may not reflect programmed settings. In a latched state, battery usage may increase, but battery status indicators will not be available. In the event that latching occurs during delivery of ATP therapy, ventricular or atrial ATP therapy delivery could continue independent of patient need. Device replacement is required if latching occurs.

Based on study of the recent third event, Guidant has determined that one of our original recommendations – programming Atrial Tachy Episode Data Storage to 0% – can cause latching in a subset of AVT devices that have previously stored atrial episode data. Guidant has determined that this newly observed latching pathway can have a significantly higher probability of occurrence (estimated at 0.086% per month) for devices that have previously stored atrial episode data. For this reason, our recommendations have been altered as follows:

REVISED Recommendations

  • Schedule a patient follow-up visit
    • As soon as possible for patients reprogrammed to 0% according to our June 20th recommendation, or any patient with Atrial Episode Data Storage programmed to less than 20%
    • Per normal scheduling if Atrial Episode Data Storage is at the nominal of 50% or is programmed to 20% or more
  • At this follow-up visit:
    1. Verify normal device function using routine clinical follow-up procedures
    2. Program Atrial Episode Data Storage to 20%
    3. Review the rate of occurrence estimates in Table 1 to evaluate the additional risk reduction benefit of programming ATP therapy OFF

Table 1 below quantifies the probability of latching associated with each of these programming options.

Important note: Atrial Episode Data Storage should not be programmed to 0% if the device has previously stored atrial episode data.

Important note: For some patients for whom atrial episode data has not been previously stored, programming to 0% may afford additional risk reduction. Contact Technical Services for additional information before programming Atrial Episode Data Storage to 0%.

Table 1. Estimated Probability of Device LatchingTable 1 footnote 1

 
Programming options Probability of latching per month Probability of latching per 6 monthsTable 1 footnote 2 Probability of latching with continuous ATP therapy per 6 monthsTable 1 footnote 2

Table 1 footnotes

Table 1 footnote 1

Probabilities are current as of July 21, 2005.

Return to table 1 footnote 1 referrer

Table 1 footnote 2

While the probabilities are stated over a six month time frame, a software solution may be available in early Q405, pending FDA approval

Return to table 1 footnote 2 referrer

Table 1 footnote 3

Although functional latching may still occur, it will not result in continuous ATP therapy

Return to table 1 footnote 3 referrer

If atrial episodes have been previously stored and Atrial Tachy Episode Data Storage is programmed to 0% [Not recommended] 0.086%
(1 per 1,160)
0.52%
(1 per 192)
0.027%
(1 per 3,700)
A) At Atrial Tachy Episode Data Storage of 50% (nominal) 0.00083%
(1 per 120,000)
0.005%
(1 per 20,000)
0.000265%
(1 per 377,000)
B) If atrial episodes have been previously stored, program Atrial Tachy Episode Data Storage to 20% 0.00043%
(1 per 232,000)
0.0026%
(1 per 38,500)
0.00014%
(1 per 714,000)
C) Program all ATP therapies to OFF No Change No Change ZeroTable 1 footnote 3
Following implementation of a software solution in early Q405Table 1 footnote 2 Zero
Return to normal programming
Zero
Return to normal programming
Zero
Return to normal programming

Devices Impacted
Only atrial therapy AVT subgroups of Guidant ICD and CRT-D product families are impacted by this anomaly. This issue does not impact standard ICDs and CRT-Ds.

Table 2. AVT Models Impacted

 
Device Family Model Numbers

Table 2 footnotes

Table 2 footnote 1

Under clinical investigation in some geographies
Devices not sold in Canada

Return to table 2 footnote ** referrer

VENTAK PRIZM AVT 1900
VITALITY AVT A135, A155
CONTAK RENEWAL 3 AVTTable 2 footnote ** M150, M155
CONTAK RENEWAL 3 AVT HETable 2 footnote ** M157, M159
CONTAK RENEWAL 4 AVTTable 2 footnote ** M170, M175
CONTAK RENEWAL 4 AVT HETable 2 footnote ** M177, M179

To date, there have been no reported failures in Canada. A list with patients implanted with affected devices will be provided to physicians with this communication if applicable.

Future Actions By Guidant
Guidant is currently developing a non-invasive software solution for VITALITY AVT and all RENEWAL AVT devices. This solution may be available in early fourth quarter, pending regulatory approval. Upon first interrogation with new programmer software, VITALITY AVT and all RENEWAL AVT devices will no longer be subject to this anomaly. At that time, memory may be re-allocated as desired for atrial episode data storage and normal programming can be resumed.

While programming options discussed above will significantly reduce the risks of functional latching in a PRIZM AVT, estimates of normal service life for remaining PRIZM AVTs indicate that few will be in service by the time a software solution can be developed and approved. Accordingly, no software update will be developed for PRIZM AVT.

Further Information
We recognize the impact of this communication on both you and your patients, and want to reassure you that patient safety remains Guidant’s primary concern. As always, if you have any questions regarding this communication, please contact:

  • Your local Guidant representative;
  • Guidant Technical Services USA at 1-800-CARDIAC (1-800-227-3422); or
  • Scott Kadwell (Guidant Canada, CRM Country Manager at 1-800-268-4487, extension 75828.

The identification, characterization, and management of medical device-related adverse incidents are dependent on the active participation of health care professionals in adverse incident reporting programmes. Any occurrences of a specific adverse incident or other serious and/or unexpected adverse incidents in patients implanted with VENTAK PRIZM AVT®, VITALITY® AVT or CONTAK RENEWAL® AVT defibrillators should be reported to Guidant Canada Corporation or Health Canada at the following addresses:

Guidant Canada Corporation
505 Apple Creek Boulevard, Unit #4
Markham, Ontario L3R 5B1
(800) 268-4487
(905) 947-5800

Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 3002C
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675

For other inquiries, please refer to contact information:
Marketed Health Products Directorate (MHPD)
Email: MHPD_DPSC@hc-sc.gc.ca
Tel: (613) 954-6522
Fax: (613) 952-7738

The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.

Sincerely,

original signed by

Scott Kadwell
Country Manager, Cardiac Rhythm Management
Guidant Canada Corporation

[Text of letter ends]
 