Notice about Health Canada advisories

Notice to Hospitals - Health Canada Issued Important Safety Information on Attachment Handles for Portable Patient Lifts

December 8, 2006

To: Hospital Chief of Medical Staff, Nursing Homes

Please distribute to Departments of Surgery, Geriatrics, Nursing, Intensive Care and/or other Departments as required and other involved professional staff and post this NOTICE in your institution.

Subject: Risk of disconnection of portable patient lifts from attachment handles, resulting in fall of patient

Attachment handles are accessories to portable patient lifts that allow the user to attach the portable lift's strap to a trolley hanging from a ceiling-mounted track (Figures 1 and 2). The design shown in Figure 2 is typical of several brands sold in Canada.

Recently, an incident involving an attachment handle occurred in which the carabiner hook at the end of the lift strap was improperly attached to the attachment handle. As a result, the lift fell on the patient, resulting in injury. Health Canada's Medical Devices Bureau (MDB) subsequently evaluated all attachment handles known to be used in Canada and discussed with affected manufacturers how potential risks could be minimized. The results of MDB's evaluations, actions taken, and recommendations are summarized in Tables 1 and 2:

Table 1: Attachment handles identified as having disconnection risks
Device Manufacturer Actions Taken & Recommendations
BHM Reacher BHM Medical Inc
Tel: 1-800-868-0441 x3407
At Health Canada's request, BHM stopped sale of its current reacher. BHM is a) providing all users with new labelling that warns of incorrect attachment methods, and b) advising users that although it believes the device to be safe when used according to instruction, Health Canada recommends that users obtain the improved version (the company will charge a fee for this).
Chiltern Wispa Attachment Handle Chiltern Invadex
Tel: 011 44 869-246470
Chiltern has agreed to stop sale and has produced a new design. Current units are being recalled and exchanged for new units at no cost to the user. Users should ensure they have exchanged their old units for the new ones.
Gaper Reacher Gaper Products Ltd. This company appears to be out of business. Users should consider replacing the Gaper with a safer design. Some of the companies mentioned in Table 2 may be of assistance. Until this replacement has occurred, users should use the Gaper carefully and consult Figures 3, 4 and 5 which show the difference between correct and incorrect connection methods for the Gaper.
Reacher Bar
(with small carabiner)
T.H.E. Medical and Just-Patient-Lifters
Tel: 1-800-565-7075
The T.H.E. reacher presents some risks if used with the original, small carabiner. The company has issued new instructions to all users and is replacing its small carabiners with newer, larger ones that are safer. Users should ensure they have exchanged their old units for the new ones.
TP Reacher Waverley Glen Systems
Tel: 1-905-850-0093
Waverley Glen has recalled the device. It is being replaced by the Griffin Reacher. The company has contacted dealers to arrange for an exchange at the user level. Users are advised to contact their Waverley Glen dealer if this exchange has not yet taken place.
Table 2: Attachment handles identified as having no disconnection risks
Device Manufacturer Status
Griffin Reacher Waverley Glen Systems Replaces the TP Reacher.
No action necessary.
Handley Attachment Handle Chiltern Invadex New design, to be used with Wispa Lite and Wispa Portable Hoists. Replaces the previous design.
Helping Hand Liko Inc.
Tel: 1-888-545-6671
Company is issuing new user instructions. Users should ensure they have received and reviewed them.
Reacher Bar
(with large carabiner)
T.H.E. Medical and Just-Patient-Lifters
Tel: 1-800-565-7075
Users should ensure they are using the large carabiner and not the smaller one. Users should make sure they have the new user manual.

Health Canada also recommends you immediately take the following actions:

  1. Review the instructions provided with your attachment handle and ensure your staff understands and follows them, especially the potential misconnection methods described in these instructions. Note however that some older user manuals do not show some or all of the possible misconnection modes; please ensure you have the latest user manual. A correct connection for some attachment handle designs is shown in Figure 2. Some of these potential misconnection modes may not pose a disconnection hazard, but others can.
  2. Ensure that all ceiling tracks are properly installed in compliance with relevant safety codes and that the lift is properly load tested. Health Canada is aware of one incident in which an improperly installed ceiling track fell from the ceiling.

Should you be aware of attachment handles for portable lifts other than those named above, please contact the Inspectorate; an assessment of these will then be conducted by MDB.

Health Canada depends on health care professionals to report adverse incidents related to medical devices. Any serious or unexpected adverse incident related to medical devices should be reported to Health Canada at the following address:

Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675

The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.

For other inquiries related to this communication, please contact Health Canada at:
Medical Devices Bureau Enquiries
Email: mdb_enquiries@hc-sc.gc.ca
Tel: (613) 957-4786
Fax: (613) 957-7318