· Published by Health Canada · Official notice
This is duplicated text of a letter from LifeScan Canada Ltd.
Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
April 10, 2006
Subject: URGENT: PRODUCT NOTIFICATION - OneTouch® SureStep® Meters with Serial Numbers from L 3152 Rx xxxxx to L 5348 Rx xxxxx Require Replacement
**PLEASE CIRCULATE TO ALL HEALTHCARE PROFESSIONALS**
Dear Healthcare Professional:
LifeScan is replacing specific OneTouch® SureStep® Meters due to a possible issue with the meter's display. If you use the OneTouch SureStep Blood Glucose Meter in your practice, or have patients who use this meter, please read the following important information to determine if meters in your practice are affected, how to receive appropriate replacement meters at no charge, and how to advise your patients of this recall.
Certain OneTouch SureStep Meters are More Likely to Develop Display Problems
The OneTouch SureStep Meter has a display that shows test results and other important information. This display is made up of "segments" that are turned on and off to form letters and numbers.
Over time, the segments in any Liquid Crystal Display (LCD) can fail. For this reason, all OneTouch SureStep Meters have a built-in display check and come with an owner's booklet that contains instructions on how to check the display when turning the meter on.
Recently, LifeScan has found that segments in the displays of certain OneTouch SureStep Meters may fail at a rate that is higher than expected. This failure can cause certain letters or numbers in the display to appear incomplete or missing. If the display is failing, it is possible that it will show a test result or a calibration code that you could misinterpret or that you can't read at all.
What You Should Do
1. Check the serial number(s) of your meter(s).
OneTouch SureStep Meters with serial numbers from L 3152 Rx xxxxx to L 5348 Rx xxxxx are more likely to have a problem with segments in the display. Please check the serial number on the back of each meter immediately to find out if it is in this range (see Figure 1 on how to check the serial number). You can also call the LifeScan Customer Care Line for assistance at 1 800 663-5521 to find out if any of your meters are affected (please have your meter with you when you call).
Figure 1

2. Call LifeScan if your meter's serial number is in the affected range.
LifeScan will replace all OneTouch SureStep Meters with serial numbers from L 3152 Rx xxxxx to L 5348 Rx xxxxx. If you have any OneTouch SureStep Meters with a serial number in this range, please call the LifeScan Customer Care Line immediately at 1 800 663-5521 to receive a replacement meter. Please have your meter with you when you call. The LifeScan Customer Care Line is open Monday to Friday from 9am to 8pm EST.
Until your replacement meter arrives, you should continue to use your current meter, as long as the display is working properly.
3. Check your meter's display before every test
If you have a meter(s) that is not in the affected range, it is still very important that you check the display before each test (see the instructions below). If you ever find that the display is not working properly, do not use the meter and call the LifeScan Customer Care Line immediately at 1 800 663-5521 for assistance.
Urge Your Patients Who Use the OneTouch SureStep Meter to Call LifeScan
While LifeScan is conducting a broad program to contact those patients identified in our records as OneTouch SureStep Meter users, please help us notify your patients of this issue. Please urge all OneTouch SureStep Meter users with affected meters to contact the LifeScan Customer Care Line at 1 800 663-5521 so we can provide them with an appropriate replacement meter at no charge. While waiting for their replacement meter, they are encouraged by LifeScan to continue to test with their current meter as long as its display is working properly.
How To Check Your OneTouch SureStep Meter's Display
To check that all display symbols are working, press and release the blue power button. The whole display will appear for 3 seconds when you turn on the meter. You should carefully look at the display to make sure all "segments", or parts of the display, turn on before each test (see Figure 2).
A similar picture of a properly working display test can also be found in your OneTouch SureStep Meter Owner's Booklet.
Figure 2
Keep This Letter And Be Sure to Contact LifeScan
Again, if you have any OneTouch SureStep Meters with serial numbers from L 3152 Rx xxxxx to L 5348 Rx xxxxx in your practice, please call the LifeScan Customer Care Line at 1 800 663-5521 to receive a replacement meter at no charge. We recommend that you keep this letter with your OneTouch SureStep Meter as a reference so that you can check your meter's display before each test.
LifeScan Canada Ltd.
#300-4170 Still Creek Drive,
Burnaby, B.C. V5C 6C6
Tel: 1 800 663-5521
www.OneTouch.ca
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
E-mail: mhpd_dpsc@hc-sc.gc.ca
Telephone: (613) 954-6522
Fax: (613) 952-7738
We apologize for any inconvenience this issue may cause, and we thank you for your continued support of LifeScan.
Sincerely,
original signed by
Les Klajnscek
President
LifeScan Canada Ltd.
Trademarks owned by Johnson & Johnson and used under license.
© 2006 LifeScan Canada Ltd. AW 086-530 03/06