This is duplicated text of a letter from LifeScan Canada Ltd.

Contact the company for a copy of any references, attachments or enclosures.

Notice about Health Canada advisories

Notice to Hospitals - Health Canada Endorsed Important Safety Information on the recall of certain lots of SureStep Blood Glucose Meters

April 18, 2006

Subject: URGENT: MEDICAL DEVICE RECALL - SureStep® Hospital Meters with Serial Numbers from L 3152 Rx xxxxx to L 5348 Rx xxxxx To Be Replaced

Dear LifeScan Customer:

LifeScan is replacing specific SureStep® Hospital Meters due to a possible issue with the meter's display. If you use the SureStep® Hospital Blood Glucose Meter in your hospital or facility, please read the following important information so that you can determine if your meter is affected and how to receive a replacement meter.

Please be aware that this recall does not affect the OneTouch®II Hospital, SureStep® Pro or SureStep® Flexx Meters, nor does it affect any test strips from LifeScan. It only affects the display of certain SureStep® Hospital Meters. (See Figure 1)

Figure 1

Meters

Certain SureStep® Hospital Meters Are More Likely To Develop Display Problems

The SureStep® Hospital Meter has a display that shows test results and other important information. This display is made up of "segments" that are turned on and off to form letters and numbers. Recently, LifeScan has found that certain SureStep® Hospital Meters may have a higher-than-expected failure rate, which can cause certain letters or numbers in the meter's display to appear incomplete or missing. If the meter's display is failing, there is a possibility that it will present a test result or a calibration code that could be misinterpreted or be unrecognizable.

Over time, the segments in any Liquid Crystal Display (LCD) can fail. For this reason, all SureStep® Hospital Meters have a built-in segment display check and come with an Operator's Guide that contains instructions on how to check the display when turning the meter on.

It is extremely important that users verify that the display is working properly before each blood glucose test.

LifeScan will replace all SureStep® Hospital Meters with serial numbers from L 3152 Rx xxxxx to L 5348 Rx xxxxx, as they are more likely to have a problem with segments in the display. If you have one or more SureStep® Hospital Meters in your facility, please follow the instructions below.

What You Should Do

  1. Immediately check all SureStep® Hospital meters in your facility for the following:
    1. Check the serial numbers on the back of the meters in your facility to find out if any of them are within this range (from L 3152 Rx xxxxx to L 5348 Rx xxxxx) as they are more likely to have a problem with segments in the display. (See Figure 2 on how to check the serial number.)
      Checking the serial number on the back of the SureStep Hospital Meter

      Figure 2

      How to check the serial number
    2. Complete the attached Fax Back Confirmation Form. Record the total number of meters that you have in your facility, then record the number of meters that you have within this range (from L 3152 Rx xxxxx to L 5348 Rx xxxxx).
    3. Fax the completed confirmation form to 1 877-292-1619. Once your Fax Back Confirmation Form is received, a LifeScan representative will contact you to schedule a replacement.
    4. Check for missing segments on all SureStep® Hospital Meters. If you determine that a meter's display is not working properly, do not use the meter and call the LifeScan Professional Line immediately at 1 888 353-0800.
  2. Ensure all staff are trained on how to check for missing segments each time they use a SureStep® Hospital meter, as specified in the Operator's Guide.
  3. While you wait for your replacement meter(s), you should continue to test blood glucose with the meter(s) you have, provided that its display is functioning properly.
  4. Be sure that you have a backup meter or an alternate means of testing glucose in your facility, in the event that the display of your SureStep® Hospital Meter(s) is not working properly. If you do not have an alternate means of testing, please call the LifeScan Professional Line at 1 888 353-0800 for assistance.

How To Check The SureStep® Hospital Meter's Display

To check that all display symbols are working, press and release the blue power button. The entire display will appear for 3 seconds when you turn on the meter. You should carefully observe the display to make sure all "segments", or parts, of the display turn on each time you test blood glucose with the meter (see Figure 3). A similar diagram of a properly functioning display test can also be found in the SureStep® Hospital Meter Operator's Guide.

Figure 3

How To Check The SureStep® Hospital Meter's Display

Keep This Letter And Be Sure to Contact LifeScan

Complete the Fax Back Confirmation Form and fax to LifeScan at 1 877-292-1619. Please call LifeScan at 1 888 353-0800 if a meter's display is not working properly, or if you do not have an alternative method of testing glucose.

LifeScan Canada Ltd.
#300-4170 Still Creek Drive,
Burnaby, B.C. V5C 6C6
Tel: 1 888 353-0800
www.OneTouch.ca

Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675

The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.

For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
E-mail: mhpd_dpsc@hc-sc.gc.ca
Telephone: (613) 954-6522
Fax: (613) 952-7738

We recommend that you keep a copy of this letter with each SureStep® Hospital Meter so that users can check the display of the meter before each test.

We apologize for any inconvenience this issue may cause, and we thank you for your continued support of LifeScan.

original signed by

Les Klajnscek
President
LifeScan Canada Ltd.


URGENT! Please Complete and Fax to: 1 877-292-1619

Attn: LifeScan Hospital Technical Support

From:
Phone:
Fax:

SureStep® Hospital Fax-Back Confirmation Form

1. Locate all SureStep® Hospital meters that you have in your facility.

  1. Record the total number of meters in the table below.
  2. Check the serial numbers on the back of the meters to find out if any of them are within this range (from L 3152 Rx xxxxx to L 5348 Rx xxxxx). Record in the table below.

Quantity of meters
Total Number of SureStep® Hospital meters in your facility
Number of SureStep® Hospital Meters with serial numbers from L 3152 Rx xxxxx to L 5348 Rx xxxxx

If you have indicated that you have SureStep® Hospital meters within the serial number range noted above, a LifeScan representative will contact you to schedule a replacement of the product.

2. Complete and sign the customer account information below.

Customer Account Number:
Institution / Facility Name:
Institution / Facility Address:
Customer Name / Title:
Signature:
Date:

3. Fax your completed form to LifeScan Customer Care at 1 877-292-1619.

If you have any questions regarding your SureStep® Hospital meters or this notification, please call the LifeScan Professional Line at 1-888-353-0800.

Trademarks owned by Johnson & Johnson and used under license.
© 2006 LifeScan Canada Ltd. AW086-500 03/06