· Published by Health Canada · Official notice
This is duplicated text of a letter from Teleflex Medical.
Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
September 2, 2008
Subject: Urgent Medical Device Recall - First Notice
Dear Valued Teleflex Medical Customer:
Teleflex Medical has issued a voluntary recall for Weck brand ligating clips including Hem-O-Lok, Horizon, Hemoclip Traditional and Hemoclip Plus products because holes were detected in the thermoformed tray of certain sterile packaged units meaning the sterile integrity cannot be guaranteed.
Based on Teleflex Medical product inspections, we have identified 317 breaches in approximately 1,478,631 packages inspected (frequency of occurrence = 0.0214 percent). The scope of the recall includes catalog numbers which are identified in the attached table that were manufactured between January 2006 through July 2008. Product with a manufacturing date of August 2008 and beyond is not a part of this recall. Health Canada has been notified of this action.
Products affected by this recall:
Attachment A provides a listing of all identified product codes and description.
For a listing of all identified product codes with their associated lot numbers that are affected by this recall, please contact Pulling Weck Canada (Customer Service 1-800-387-9699).
Our records indicate that the product included in the scope of this recall has been distributed in Canada. Teleflex Medical is aware that Health Canada will post this recall letter on the Health Canada website. In order to provide the highest level of quality product to our customers, we are notifying our customers to take the following actions:
We apologize for any inconvenience this may have caused. If you require additional information or clarification regarding this matter, please contact Customer Service at 1-800-387-9699 ; 905- 943-9000, or your Local Sales Representative at the following numbers:
PILLING WECK CANADA is in the process of inspecting all of our Canadian Inventory. 100% of this inventory will be inspected before it is released for sale. The inspected product will be marked with a YELLOW STICKER on each Cartridge and on each BOX. This clearly identifies product that has been inspected for holes in the hard plastic tray that encloses these affected ligation clips. If you have any of this marked product in inventory you need not return it.
Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-marketing adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any cases of serious or unexpected adverse incidents in patients using Weck Brand Ligating Clips should be reported to Teleflex Medical or Health Canada at the following addresses:
Teleflex Medical
Pilling Weck Canada
1-165 Gibson Drive
Markham (Ontario) L3R 3K7
Telephone: 800-387-9699
Fax: 905-943-9001
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
E-mail: MHPD_DPSC@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738
HORIZON
HEM-O-LOK
HEMOCLIP AUTO
HEMOCLIP TRADITIONAL
HEMOCLIP PLUS
This HTML document is not a form. Its purpose is to display the information as found on the form for viewing purposes only. If you wish to use the form, you must use the alternate format below.
Please check the appropriate box and return this form by FAX to 905-943-9001 immediately:
Return all affected product to:
PILLING WECK CANADA
Recalled Product
Attention: Recall Department
1-165 Gibson Drive, Markham
Ontario, L3R 3K7
If you received the product from a distributor, please state from which distributor.
Name:
Location:
Address:
City:
Province:
(Hospital Name)
(Date)
Print Name
(Signature)
(Telephone Number)
(Email Address)