Health ProductsDear Healthcare Professional Letter

Urgent Medical Device Correction: Liko Uno Patient Lift - For Health Professionals

· Published by Health Canada · Official notice

Details

Alert type: Dear Healthcare Professional Letter
Category: Medical Device
Audience: Healthcare Professionals
Department: Health Canada
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Images

Details

This is duplicated text of a letter from Hill-Rom.
Contact the company for a copy of any references, attachments or enclosures.

Notice about Health Canada advisories

Health Canada Endorsed Important Safety Information on Liko Uno Patient Lift

November 11, 2008

Subject: Urgent Medical Device Correction: Liko Uno Patient Lift

FSCA-identifier: Mod 403
Type of action: Device Modification

To: Facility Risk Manager/Facility Administrator

Affected Devices:

MODELS:

  • Uno 100 EM, Uno 100 EE, Uno 101, Uno 102, Uno 102 EM, Uno 102 EE, and Uno 102 ES

RANGE OF SERIAL NUMBERS AFFECTED:

  • Uno 100: 7090001-7096200
  • Uno 101: 10001-11000
  • Old Uno 102: 20001-21300
  • New Uno 102: 30001-48100

PRODUCTION DATES:

  • Manufactured prior to June 26, 2008

Background:

Liko, a Hill-Rom company, has initiated a field correction program for the Liko Uno Patient Lift, applicable only to the models and serial numbers listed above.

The actuator of the lift has the potential to separate when the upper arm is fully extended (see photo). If this happens when a patient is being lifted the lift will fail, resulting in a patient fall.

photograph of patient lift indicating the point of separation

To date, Liko has received 18 reports of actuator failures, one report of a death from complications associated with a hip fracture, and one report of a fractured femur.

Action to be taken:

Immediately take the unit out of service and contact your authorized Liko distributor or Hill-Rom to discuss available options.

Transmission of this Field Safety Notice:

Please forward a copy of this letter to any other facility personnel you deem appropriate.

If you have any questions concerning this request, please contact Hill-Rom Technical Support using the attached listing for your geographic region.

Hill-Rom
Telephone: 1-800-267-2337
E-mail: jeff.moorehead@hill-rom.com
Distributor Peridot Medical
Telephone: 905-541-1798
E-mail: thuxley@sympatico.ca

Any suspected adverse reaction can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675

The Medical Devices Problem Report Form and the Guidelines can be found on the Health Canada web site.

For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
E-mail: MHPD_DPSC@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738

Regards,

Hill-Rom Technical Support

Technical Support Contact Information

Important: This is a comprehensive list of Technical Support contact information for Hill Rom offices worldwide. If this is attached to another document or letter, it does not mean that every country listed below is affected by that document.

All other countries for support of products manufactured in the US:
Telephone: 812-934-1348 (8 am - 5 pm ET (New York time))
e-mail: gary.lakes@hill-rom.com

All other countries for support of products manufactured outside the US:
Telephone: +33 (0)2 97 50 92 45
e-mail: leny.rodriguez@hill-rom.com