November 11, 2008
Subject: Urgent Medical Device Correction: Liko Uno Patient Lift
FSCA-identifier: Mod 403
Type of action: Device Modification
To: Facility Risk Manager/Facility Administrator
Affected Devices:
MODELS:
RANGE OF SERIAL NUMBERS AFFECTED:
PRODUCTION DATES:
Background:
Liko, a Hill-Rom company, has initiated a field correction program for the Liko Uno Patient Lift, applicable only to the models and serial numbers listed above.
The actuator of the lift has the potential to separate when the upper arm is fully extended (see photo). If this happens when a patient is being lifted the lift will fail, resulting in a patient fall.

To date, Liko has received 18 reports of actuator failures, one report of a death from complications associated with a hip fracture, and one report of a fractured femur.
Action to be taken:
Immediately take the unit out of service and contact your authorized Liko distributor or Hill-Rom to discuss available options.
Transmission of this Field Safety Notice:
Please forward a copy of this letter to any other facility personnel you deem appropriate.
If you have any questions concerning this request, please contact Hill-Rom Technical Support using the attached listing for your geographic region.
Hill-Rom
Telephone: 1-800-267-2337
E-mail: jeff.moorehead@hill-rom.com
Distributor Peridot Medical
Telephone: 905-541-1798
E-mail: thuxley@sympatico.ca
Any suspected adverse reaction can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and the Guidelines can be found on the Health Canada web site.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
E-mail: MHPD_DPSC@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738
Regards,
Hill-Rom Technical Support
Technical Support Contact Information
Important: This is a comprehensive list of Technical Support contact information for Hill Rom offices worldwide. If this is attached to another document or letter, it does not mean that every country listed below is affected by that document.
All other countries for support of products manufactured in the US:
Telephone: 812-934-1348 (8 am - 5 pm ET (New York time))
e-mail: gary.lakes@hill-rom.com
All other countries for support of products manufactured outside the US:
Telephone: +33 (0)2 97 50 92 45
e-mail: leny.rodriguez@hill-rom.com