March 9, 2011
Dear Risk Manager/Facilities Administrator/Biomed Engineer:
Please distribute to relevant Departments (Biomedical Engineering, Surgery, Emergency Medicine, Paediatrics, Anaesthesia, Geriatrics, Internal Medicine, Nursing, Intensive Care and/or other Departments as required), and other involved professional staff and post this Notice in your institution.
Subject: Recall of Medical Bed Model FLH668NDCM with potential for head or foot section to drop unexpectedly
Stryker Medical would like to notify you of a potential situation regarding the "screw box" on the Bertec Model FLH668NDCM beds for which a relevant Canadian incident was reported to Health Canada. While no injuries have occurred, we have identified the potential for either the head section or the foot section of the bed to drop unexpectedly, if preventive maintenance has not been conducted, as specified in the owner's manual. The owner's manual included instructions for the annual cleaning and lubrication of the screw box. The potentially affected units were manufactured more than 10 years ago. While the units are currently beyond their intended life, we are taking this opportunity to notify users, as they may still be in-use.
Immediate Action Required:
Should you have any further questions or concerns, please do not hesitate to contact Stryker Medical at 1-905-690-5700 ext. 2198, or via e-mail at CARAQA@stryker.com.
Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-market adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any cases of serious or unexpected adverse incidents related to FLH668NDCM medical beds should be reported to Stryker Canada or Health Canada at the following addresses:
Stryker Canada
45 Innovation Drive
Hamilton, Ontario
L9H 7L8
Toll: 1-800-668-8324
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada Web site.
For other medical device inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate
E-mail: mhpd_dpsc@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738
To change your mailing address or fax number, contact the Market Authorization Holder (Industry).
original signed by
Nicole Khanna
Quality Assurance & Post Market Vigilance
March 9, 2011
Internal Reference# RA 2011-044 [Correction Date: March 9, 2011]
Description: Medical Bed Model FLH668NDCM
Hospital:
Please complete and return this form immediately to fax# 1-905-690-5718 or email to CARAQA@stryker.com.
I have received the notification from Stryker dated March 9, 2011 stating that they have initiated a product correction for the above referenced product.
COMMENTS:
(Signature) Risk Manager/Facilities Administrator/Biomed Engineer:
Date:
Hospital (Print):
Date: