This is duplicated text of a letter from Covidien Canada.
Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
January 13, 2012
Attention: Risk Management Director and O.R Materials Management
Please forward this communication to all surgeons/surgical personnel.
Subject: Urgent Medical Device Recall, Contraindication For Thoracic Use, Duet TRS Loading Units
Dear Customer,
Covidien (formerly United States Surgical, a division of Tyco Healthcare Group, L. P.) is conducting a voluntary Recall of all production lots for the Duet TRS™ Universal Straight and Articulating Single Use Loading Units (SULU) with respect to the use of this product family in the thoracic cavity.
Serious injuries and deaths have occurred due to the failure mode associated with this Recall. Covidien has received reports of 3 deaths and 13 serious injuries following the application of Duet TRS™ Loading Units in the thoracic cavity. We have concluded that Duet TRS™ has the potential to injure adjacent anatomical structures within the thorax which may result in life threatening postoperative complications.
Users of Duet TRS™ Loading Units should not use the device in thoracic surgery in both adult and pediatric populations. Users of Duet TRS™ may continue to use Duet TRS™ in other applications. Covidien will be revising the Instructions for Use for the Duet TRS™ to contraindicate the Duet TRS™ for use in thoracic procedures in adult and pediatric populations. In the interest of patient safety we are initiating this Recall in advance of that change.
This Recall only applies to the Duet TRS™ loading units and only for the use of this product in the thoracic cavity. Other Endo GIA™ families of SULUs are not affected.
The affected product codes and descriptions are as follows:
DUET4535
Duet TRS™ 45 3.5MM
STRAIGHT SULU
DUET4535A
Duet TRS™ 45 3.5MM
ARTICULATING SULU
DUET4548
Duet TRS™ 45 4.8MM
STRAIGHT SULU
DUET4548A
Duet TRS™ 45 4.8MM
ARTICULATING SULU
DUET6035
Duet TRS™ 60 3.5MM
STRAIGHT SULU
DUET6035A
Duet TRS™ 60 3.5MM
ARTICULATING SULU
DUET6048
Duet TRS™ 60 4.8MM
STRAIGHT SULU
DUET6048A
Duet TRS™ 60 4.8MM
ARTICULATING SULU
Required actions:
This action is being taken with the knowledge of Health Canada. If you have any questions or concerns, please do not hesitate to contact your Covidien representative or Covidien Customer Service at 1-877-664-8926, option 1.
Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-market adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any cases of serious injury or other serious or unexpected adverse incidents for medical devices in patients using Duet TRS™ Single Use Loading Units should be reported to Covidien Canada or Health Canada at the following addresses:
Joseph DiMarzo
Director, Regulatory Affairs and Quality Assurance
Covidien Canada
7300 Trans Canada Highway,
Pointe Claire, Quebec H9R 1C7
Telephone : 1-877-664-8926
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
Health Canada
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada Web site.
For other medical device inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate
E-mail: mhpd_dpsc@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738
To change your mailing address or fax number, contact the Market Authorization Holder (Industry).
Sincerely,
original signed by
Joseph Di Marzo
Director, Regulatory Affairs and Quality Assurance
Covidien Canada
This HTML document is not a form. Its purpose is to display the information as found on the form for viewing purposes only. If you wish to use the form, you must use the alternate format below.
Help on accessing alternative formats, such as Portable Document Format (PDF), Microsoft Word and PowerPoint (PPT) files, can be obtained in the alternate format help section.
DUET TRS™ Loading Unit
Contraindication for Thoracic Use
CUSTOMER RESPONSE FORM # 1
Acknowledgement with no product to return
AS USAGE IS INTENDED FOR NON-THORACIC PROCEDURES
The following information must be completed and faxed to:
Covidien Canada,
Dawn Boyce - RA/QA Dept.
Fax no: 514-695-2379
Email: dawn.boyce@covidien.com
Date:
Name of Person Completing this form:
Title:
Direct Phone:
Email:
Account Name:
Covidien Account #:
Address:
City:
Prov:
PCode:
How did the account purchase this product? (Please check)
Direct from Covidien
From a Distributor
Distributor Name:
Address:
City:
Prov:
PCode:
(please sign).
I acknowledge receipt of the duet trs™ Loading unit recall notification dated january 13, 2012 and will not be returning inventory.
DUET TRS™ Loading Unit
Contraindication for Thoracic Use
Customer Response Form # 2 - Product to return
Product intended for thoracic use must be returned.
The following information must be completed and faxed to:
Covidien Canada,
Dawn Boyce - RA/QA Dept.
Fax no: 514-695-2379
Email: dawn.boyce@covidien.com
We have the following affected product in stock and understand that Covidien Customer Service will contact with me with a Return Goods Authorization number: (Covidien Customer Service can be reached at 1-877-664-8926, option 1).
| Item number | Lot number | Quantity found (eaches) | Customer po # |
|---|---|---|---|
Hospital Name
Address
City, Province
Contact Name (Please Print)
Title / Department
Signature
Date
Phone Number
Fax Number