· Published by Health Canada · Official notice
This is duplicated text of a letter from Hospira Healthcare Corporation. Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
March 9, 2012
Dear Health Care Professional:
Subject: Urgent-device field correction GEMSTAR™ Bolus Cord
List Number 13027
Hospira, Inc. has received customer reports of bolus delivery failures when using the GemStar bolus cord. Possible failures may be a result of cracked or damaged bolus cords which may result in an open or short circuit of the bolus cord.
When possible, Hospira recommends the use of the bolus push button on the pump. Please reference the pictures below highlighting the Bolus Button.

Bolus button located on the Top of the Pump
Hospira's review has determined that the likely root cause is associated with the force applied to remove the bolus cord from the base of the GemStar unit. If excessive force is used to remove the bolus cord, it may result in broken or frayed wiring at the base of the cord.
Care should be taken when removing the bolus cord from the base of the GemStar unit. Proper technique for removing the bolus cord is to firmly grasp the connector and pull straight out. Do not twist or bend the cord or connector. Reference the pictures below. DO NOT USE IF DAMAGED.

Grasp the connector to pull bolus cord from the base of the pump

Do not pull or bend the bolus cord below the base
You should refer to the Technical Service Manual for proper inspection of your existing cords. Below are excerpts from the service manual for your convenience (Hospira GemStar Technical Service Manual, Section Routine Maintenance, Remote Bolus Cord and Switch).
Please report all issues related to the use of the bolus cord to Hospira, using the information provided below. As stated in the GemStar Technical Services Manual, you should discontinue use of any cracked or damaged cords and call Hospira for an immediate replacement. Please continue to monitor the bolus cords on a regular basis.
| Hospira Contact | Contact Information | Areas of Support |
|---|---|---|
| Hospira Complaint & Safety Department | 1-866-488-6088 Option 6 | To report adverse events or product complaints |
| Hospira Canada Service Center | 1-866-488-6088 Option 5 | Technical Assistance |
Hospira is exploring ways to make the bolus cord less vulnerable to damage by a re-design or other methods. Once our investigation is completed, Hospira will implement any necessary improvements.
Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-marketing adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any case of serious delay in critical therapy or other serious or unexpected adverse incidents for medical devices in patients using GemStar Pump Sets should be reported to Hospira Healthcare Corporation or Health Canada at the following addresses:
Hospira Healthcare Corporation
Complaint and Safety Department
1111 Dr Frederik- Philips, Suite 600
Saint-Laurent (Québec), H4M 2X6
To Report Adverse Reaction, Consumers and Health Professionals may call toll free:
Tel: 866-488-6088 Option 6
Fax: 877-906-0208
ProductcomplaintsCA@hospira.com
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada Web site.
For other medical device inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate
E-mail: mhpd_dpsc.public@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738
To change your mailing address or fax number, contact the Market Authorization Holder (Industry).
Sincerely,
original signed by
Beryl Chan
Regional Director, Quality and Regulatory Affairs
Hospira Healthcare Corporation