Health ProductsDear Healthcare Professional Letter

GemStar Bolus Cord - Urgent - Potential Risk Due to Device Failure - For Health Professionals

· Published by Health Canada · Official notice

Details

Alert type: Dear Healthcare Professional Letter
Category: Medical Device
Audience: Healthcare Professionals
Department: Health Canada
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Images

Details

This is duplicated text of a letter from Hospira Healthcare Corporation. Contact the company for a copy of any references, attachments or enclosures.

Notice about Health Canada advisories

Health Canada Endorsed Important Safety Information on GemStar Bolus Cord

March 9, 2012

Dear Health Care Professional:

Subject: Urgent-device field correction GEMSTAR™ Bolus Cord
List Number 13027

Hospira, Inc. has received customer reports of bolus delivery failures when using the GemStar bolus cord. Possible failures may be a result of cracked or damaged bolus cords which may result in an open or short circuit of the bolus cord.

  • Potential risk due to failure of the bolus cord is the potential for an unrequested bolus delivery or the failure to deliver requested medication to the patient.
  • The amount of unrequested delivery will be limited by programming lockout volumes.
  • An unrequested narcotic delivery may cause over sedation requiring possible medical intervention, but is unlikely to result in serious or life threatening injuries. 
  • The failure to deliver requested narcotic medication may result in the lack of adequate pain relief.

When possible, Hospira recommends the use of the bolus push button on the pump. Please reference the pictures below highlighting the Bolus Button.

Bolus button located on the Top of the Pump

Hospira's review has determined that the likely root cause is associated with the force applied to remove the bolus cord from the base of the GemStar unit. If excessive force is used to remove the bolus cord, it may result in broken or frayed wiring at the base of the cord.

Care should be taken when removing the bolus cord from the base of the GemStar unit. Proper technique for removing the bolus cord is to firmly grasp the connector and pull straight out. Do not twist or bend the cord or connector. Reference the pictures below. DO NOT USE IF DAMAGED.

Grasp the connector to pull bolus cord from the base of the pump

Do not pull or bend the bolus cord below the base

You should refer to the Technical Service Manual for proper inspection of your existing cords. Below are excerpts from the service manual for your convenience (Hospira GemStar Technical Service Manual, Section Routine Maintenance, Remote Bolus Cord and Switch).

  1. Verify the bolus cord is free from fraying and torn insulation.
  2. Verify the switch body is free of damage and evidence of fluid ingress.
  3. Verify the bolus port connector pins are present and in good condition.
  4. Press the bolus button and verify the continuity across pins 1 and 2.
  5. Connect the bolus cord to the bolus port on the bottom of the infuser. Verify the bolus cord plug is properly engaged/ retained.

Please report all issues related to the use of the bolus cord to Hospira, using the information provided below. As stated in the GemStar Technical Services Manual, you should discontinue use of any cracked or damaged cords and call Hospira for an immediate replacement. Please continue to monitor the bolus cords on a regular basis.

Hospira Contact Information
Hospira Contact Contact Information Areas of Support
Hospira Complaint & Safety Department 1-866-488-6088 Option 6 To report adverse events or product complaints
Hospira Canada Service Center 1-866-488-6088 Option 5 Technical Assistance

Hospira is exploring ways to make the bolus cord less vulnerable to damage by a re-design or other methods.  Once our investigation is completed, Hospira will implement any necessary improvements.

Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-marketing adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any case of serious delay in critical therapy or other serious or unexpected adverse incidents for medical devices in patients using GemStar Pump Sets should be reported to Hospira Healthcare Corporation or Health Canada at the following addresses:

Hospira Healthcare Corporation
Complaint and Safety Department
1111 Dr Frederik- Philips, Suite 600
Saint-Laurent (Québec), H4M 2X6
To Report Adverse Reaction, Consumers and Health Professionals may call toll free:
Tel: 866-488-6088 Option 6
Fax: 877-906-0208
ProductcomplaintsCA@hospira.com

Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675

The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada Web site.

For other medical device inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate
E-mail: mhpd_dpsc.public@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738

To change your mailing address or fax number, contact the Market Authorization Holder (Industry).

Sincerely,

original signed by

Beryl Chan
Regional Director, Quality and Regulatory Affairs
Hospira Healthcare Corporation