Health ProductsDear Healthcare Professional Letter

Archived - Non-sterile medical devices labeled as sterile manufactured by A & A Medical

· Published by Health Canada · Official notice

Details

Alert type: Dear Healthcare Professional Letter
Category: Medical Device
Audience: Healthcare Professionals
Department: Health Canada
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Details

This is duplicated text of a letter issued by the Therapeutic Products Directorate, Medical Devices Bureau.
Contact the company for a copy of any references, attachments or enclosures.

Notice about Health Canada advisories

[Text of letter begins]

Important Medical Devices Safety Advisory

March 22, 2002

Non-sterile medical devices labeled as sterile manufactured by A & A Medical, Rocket USA and LifeQuest

To:

Hospital Administrators
Director, Department of Obstetrics and Gynecology
Director, Central Supply
Provincial Medical Officers of Health
Ambulatory Obstetric/Gynecology Clinics
Society of Obstetricians and Gynecologists of Canada

Dear Health Professional(s),

This is to inform you of an important medical device problem with all products labeled as sterile or as ethylene oxide processed manufactured by A & A Medical of Alpharetta, GA. These devices may also be sold under the trade names of "A & A Medical", "Rocket USA" and "LifeQuest".

Some products manufactured by this company have been labeled and sold as sterile or as ethylene oxide processed but in fact may not have undergone any sterilization process. Infections associated with the use of these devices have been reported in the U.S.A. To date, Health Canada has not received reports of infections related to the use of these devices in Canada. Although the only Rocket USA product licensed for sale in Canada is the Rocket Aspiration Set with Locking Syringe, other products have been sold here.

Recommendations:

  1. Do not use any devices labeled as sterile or as ethylene oxide processed made by "A & A Medical", "Rocket USA", or "LifeQuest".
  2. If you have any products labeled as sterile or as ethylene oxide processed made by these companies, place them in quarantine, contact Health Canada immediately and inform the distributor who sold you the device(s).

Health Canada is working to identify all distributors of these products, and is attempting to have them advise all their customers who received the affected products.

Background Information:

A recall has been initiated in the USA as posted on the FDA website of all products manufactured under the name of "A & A Medical", "Rocket USA", or "Lifequest" that are labeled as sterile or as ethylene oxide processed. This firm manufactures many types of Ob/Gyn and surgical devices. The recall includes, but is not limited to curettes (flexible and rigid), uterine dilators, endometrial sampling sets, fetal blood samplers, fetal bladder drains, laparoscopy accessories, bone marrow needles, aspiration sets, biopsy pipettes, and harvesting pumps used in in vitro fertilization. The recall applies to all products labeled as sterile that have been shipped since 1999.

For additional information on the US recall, you may visit the US FDA website at: http://www.fda.gov/cdrh/recalls/recall31402.html

Contacts:

Individuals seeking additional information should leave a message on the Medical Devices Hotline at 1-800-267-9675. A Health Canada official will return your call.

Should you wish to submit information concerning this issue, please send it to:

Health Products and Food Branch Inspectorate
National Coordination Centre
11 Holland Avenue, AL: 3002C
Holland Cross Building,
Tower A, 2nd Floor
Ottawa, Ontario,
K1A 0K9,
Telephone: 1-800-267-9675,
Facsimile: (613) 954-0941

original signed by

Don Boyer A/Director,
Medical Devices Bureau
don_boyer@hc-sc.gc.ca
Tel: 613 957-4786

Any suspected adverse incident can also be reported to:

Health Products & Food Branch Inspectorate
Health Canada
Address Locator: 3002C
Ottawa, Ontario,
K1A 0K9
Tel: The Medical Devices Hotline 1-800-267-9675

[Text of letter ends]
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime