Philips IntelliVue Patient Monitors - Important Safety Information - Notice to Hospitals

· Published by Health Canada · Official notice

Details

Alert type: Notice to Hospitals
Category: Medical Device
Audience: Healthcare Professionals
Department: Health Canada
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Introduction

This is duplicated text of a letter from Philips Healthcare.
Contact the company for a copy of any references, attachments or enclosures.

Notice about Health Canada advisories

Details

Notice to Hospitals - Health Canada Endorsed Important Safety Information on Philips IntelliVue Patient Monitors

June 1, 2012

To: Hospital Chief of Medical Staff

Please distribute to all relevant Departments [Intensive Care Unit, Emergency Medicine, Paediatric unit, and/or other Departments as required], and other involved professional staff and post this Notice in your institution.

Subject: Important safety information - Philips IntelliVue Patient Monitors IntelliVue model (product code): MP20 (M8001A), MP30 (M8002A), MP50 (M8004A), MP60 (M8005A), MP70 (M8007A), MP80 (M8008A), MP90 (M8010A), MX700 (865241), MX800 (865240).

Philips Healthcare, in collaboration with Health Canada, is providing this letter to inform you of a potential safety issue that may affect the IntelliVue Patient Monitors. Philips has recently received reports of patient vital signs monitors not alarming (either visual or audible) at the central station. This could result in serious injury to patients requiring emergency care due to delay of treatment. The affected units include IntelliVue Patient Monitors with software (SW) Revision H (up to and including H.15.36). The issue is summarized below.

  • Under certain circumstances, alarms at the affected patient monitors are not announced (either visual or audible) at the central station. If this occurs, staff monitoring the central station may not be alerted to a patient's condition requiring intervention. This may lead to a delay in medical intervention and patient treatment.
  • The primary alarm function at the bedside monitor is not affected. All physiological information transferred from the bedside monitor is correctly displayed at the central station.
  • Philips is providing a software upgrade for all affected devices at no charge. Philips has issued a Field Safety Notice (dated February 9, 2012) to customers regarding this issue. A Philips Healthcare representative should have contacted you to arrange for the software upgrade.
  • Please follow-up so that all users are aware of this hazard. Do not rely on the alarms at the central station until the software upgrade is installed.
  • Review this information with all staff members who interface with the central station and need to be aware of the contents of this communication. Please retain a copy with the equipment Instruction for Use.

The Product Number and Serial Number is contained on the devices product label, located on the front of the device. The SW revision can be accessed via the Revision Screen at the bedside monitor.

If you need any further information or support concerning this issue, please contact your local Philips Healthcare Representative at: 1-800-567-1080.

Please complete the Medical Device Correction Confirmation Form and return by fax to Philips Canada Regulatory Dept at 905-201-4323.

Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-marketing adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any serious or unexpected adverse incidents involving IntelliVue Patient Monitors should be reported to Philips Healthcare or Health Canada at the following addresses:

Philips Healthcare
281 Hillmount Road
Markham, ON
Canada L6C 2S3
Telephone: 1-800-567-1080
Fax: 1-800-418-4285

To correct your mailing address or fax number, contact Philips Healthcare.

Any suspected adverse incident can be reported to:
Health Products and Food Branch Inspectorate
Health Canada
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline: 1-800-267-9675

The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada Web site.

For other medical device inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate
E-mail: mhpd_dpsc.public@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738

Sincerely,

original signed by

Althea Lawrence
Director of Regulatory Affairs & Quality Assurance
Philips Healthcare

This HTML document is not a form. Its purpose is to display the information as found on the form for viewing purposes only. If you wish to use the form, you must use the alternate format below.

Download PDF document

Medical Device Correction Confirmation Form

Recall No. PH12003R

Philips Healthcare
281 Hillmount Road
Markham, ON

Product: Philips IntelliVue Patient Monitors:
MP20 (M8001A), MP30 (M8002A), MP50 (M8004A), MP60 (M8005A), MP70 (M8007A), MP80 (M8008A), MP90 (M8010A), MX700 (865241), MX800 (865240)

Please complete this form and fax to:

Philips Healthcare
Attn: Canada Regulatory Dept
Fax: 905-201-4323

Facility Name:

Address:

This will confirm receipt of the Medical Device Correction letter (Recall No. PH12003R). We have read, understand, and will follow the prescribed instructions.

Signed:

Date:

Name:
(please print or type)

Title:
(please print or type)

Department:

Telephone:

Fax:

E-mail:

A division of Philips Electronics Ltd.
Telephone: (905) 201-4100 / Toll Free: 1-877-744-5633
Facsimile: (905) 201-4323 / Toll Free: 1-800-418-4285
www.healthcare.philips.com

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime