Health Products

Navios Flow Cytometry Software (two different models)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
Navios Flow Cytometery SoftwareA52103
All lots
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
Navios Flow Cytometery SoftwareA52102

Reason

When running a Multi-tube Worklist, there is a slight possibility that a Listmode file may be duplicated. When this happens, the next tube on the worklist is skipped (not sampled). Subsequent tubes on the worklist are processed correctly. The duplicated listmode file contains all of the correct data events and FCS keyword information but will have the skipped tube's Listmode filename assigned to it. Therefore any panel report generated from the skipped tube may have erroneous but credible results.