H. Riata I Passive Fixation Lead (Steroid Eluting)
1560
Not Applicable
Manufacturer
St-Jude Medical
Cardiac Rhythm Management Division
I. Riata I Active Fixation Lead (Steroid Eluting)
1590
Not Applicable
Manufacturer
St-Jude Medical
Cardiac Rhythm Management Division
F. Riata Single Coil Passive Fixation Lead (Steroid Eluting)
1572
Not Applicable
Manufacturer
St-Jude Medical
Cardiac Rhythm Management Division
G. Riata Single Coil Active Fixation Lead (Steroid Eluting)
1582
Not Applicable
Manufacturer
St-Jude Medical
Cardiac Rhythm Management Division
D. Riata Passive Fixation Tachyarrhythmia Lead (Steroid Eluting)
1570
Not Applicable
Manufacturer
St-Jude Medical
Cardiac Rhythm Management Division
E. Riata Active Fixation Tachyarrhythmia Lead (Steroid Eluting)
1580
Not Applicable
Manufacturer
St-Jude Medical
Cardiac Rhythm Management Division
B. Riata ST Active Fixation Integrated Bipolar ICD Lead
7010
Not Applicable
C. Riata Passive Fixation True Bipolar ICD Lead
7040
Not Applicable
Manufacturer
St-Jude Medical
Cardiac Rhythm Management Division
A. Riata ST Active Fixation Dual Coil LCD Lead (Steroid Eluting)
7000
Introduction
Recalled products
Riata ST Active Fixation Dual Coil LCD Lead (Steroid Eluting)
Riata ST Active Fixation Integrated Bipolar ICD Lead
Riata Passive Fixation True Bipolar ICD Lead
Riata Passive Fixation Tachyarrhythmia Lead (Steroid Eluting)
Riata Active Fixation Tachyarrhythmia Lead (Steroid Eluting)
Riata Single Coil Passive Fixation Lead (Steroid Eluting)
Riata Single Coil Active Fixation Lead (Steroid Eluting)
Riata I Passive Fixation Lead (Steroid Eluting)
Riata I Active Fixation Lead (Steroid Eluting)
Reason
Riata and Riata ST silicone leads have demonstrated a high rate of externalized conductors, i.e. interior, ethylene tetrafluoroethylene (ETFE) covered, lead conductor wires extruding from the device lumen into the intravascular space.