Health Products

Dimension Systems

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
GA2196
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
B. Dimension Xpand System - Creatinine(CREA) AssayDF33A
GA2196
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
C. Dimension EXL System - Creatinine(CREA) AssayDF33A
GA2196
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
A. Dimension Chemistry System CREA/CreatinineDF33A

Introduction

Recalled products
  1. Dimension Chemistry System CREA/Creatinine
  2. Dimension Xpand System - Creatinine (CREA) Assay
  3. Dimension EXL System - Creatinine (CREA) Assay

Reason

Siemens Healthcare diagnostics have received customer complaints regarding the potential of Dimension CREA to increase QC recovery and imprecision over time from open well. QC results shift high over time within a flex well and then recover near the QC target or expected value in a newly opened fresh well. The change in QC recovery within a flex well can be as high as 0.2mg/dl in a 24 hour period. This challenges the Dimension CREA flex open well stability claim of 5 days.