B. Dimension Xpand System - Creatinine(CREA) Assay
DF33A
GA2196
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
C. Dimension EXL System - Creatinine(CREA) Assay
DF33A
GA2196
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
A. Dimension Chemistry System CREA/Creatinine
DF33A
Introduction
Recalled products
Dimension Chemistry System CREA/Creatinine
Dimension Xpand System - Creatinine (CREA) Assay
Dimension EXL System - Creatinine (CREA) Assay
Reason
Siemens Healthcare diagnostics have received customer complaints regarding the potential of Dimension CREA to increase QC recovery and imprecision over time from open well. QC results shift high over time within a flex well and then recover near the QC target or expected value in a newly opened fresh well. The change in QC recovery within a flex well can be as high as 0.2mg/dl in a 24 hour period. This challenges the Dimension CREA flex open well stability claim of 5 days.