AW Volumeshare - Seno Advantage Software Option

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Assistive/Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
'00000193277GE1, '000004410002GS
Manufacturer
GE Medical Systems SCS
AW Volumeshare - Seno Advantage Software Option5341143-X, 5342425-X
> 10 Numbers, Contact manufacturer
Manufacturer
GE Medical Systems SCS
AW Volumeshare - Seno Advantage Software Option5183685-X, 5183859-X

Reason

GE Healthcare has become aware that patient mismatch without a warning message could occur on Seno Advantage (SA) Workstations under the following circumstances:
1. New non-mammographic images for patient a are being received (for example, via the network) at the same time that mammographic images for patient a are displayed on the high-resolution black and white monitors.
2. The non-mammographic images are then selected and displayed on the color monitor.
3. Mammographic images from a different patient are then selected on the high-resolution black and white monitors.