Health Products

Biotestcell P3

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2225011
Manufacturer
BioRad Medical Diagnostics GMBH
A. Biotestcell P3816 017

Introduction

Recalled Products

A. Biotestcell P3

Reason

The Biotestcell P3 are reagent red blood cells with polyvalent antigens of three single blood donors packaged in 3 separate vials for the detection of red cell antiobodies. The manufacturer received one complaint about an incorrectly labeled vial within the package. Normally, the vials come numbered P1, P2, P3. In this case the package contained 1 bottle of P1 and 2 bottles of P2.