Health Products

Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System, Part Number 775222, Software Revision 1.5.1.0

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Beckman Coulter Inc.
Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System, Part Number 775222, Software Revision 1.5.1.0775222

Introduction

Recalled Products:

Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System, Part Number 775222, Software Revision 1.5.1.0

Reason

The DXH SMS (Slidemaker Stainer) System is generating an "invalid code" message (error code 75) when the LIS transmits any field in the record that is not populated (nil). The LIS will retransmit the test order to the analyzer and the DXH SMS will reject the retransmission and send a message to the LIS that the test order was already received/processed.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime