Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System, Part Number 775222

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System, Part Number 775222
Manufacturer
Beckman Coulter Inc.
775222

Introduction

Recalled Products:

Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System, Part Number 775222

Reason

The DXH SMS with software version 1.5.1.0 will not read Codabar labels at the primary barcode reader when the AIM-16 checksum digit calculation is used with any leading and trailing characters that are not both "a" (e.g. a12345b). The Codabar barcode symbology uses a single leading and trailing character as start and stop indicators, with characters such as a, b, c or d. when the Codabar symbology is used the labels can contain either matching or non-matching start and stop characters. If the character a is used for both the start and stop character (i.e. a12345a) the DXH SMS will read the bar code label successfully. If any other start and stop characters are used the system will produce no read from the primary barcode reader.