Samaritan PAD SAM 300 and 300P

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
040000501 - 0700032917B. Samaritan PAD SAM 300P
Manufacturer
HeartSine Technologies, Ltd.
SAM300P
040000501 - 0700032917A. Samaritan PAD SAM 300
Manufacturer
HeartSine Technologies, Ltd.
SAM300

Introduction

Recalled Products:
  1. Samaritan PAD SAM 300
  2. Samaritan PAD SAM 300P

Reason

Devices within these serial number ranges may be affected by one of both of these 2 identified issues :

Issue one: devices may intermittently, or sometimes repeatedly, switch themselves on, and remain on for 10 minutes, which could eventually cause premature depletion of battery capacity. Should this condition not be detected the battery may become depleted below a level where it could power the device to safely deliver therapy if required.

Issue two: the battery management software in affected devices may misinterpret changes in voltage which occur during the normal operation of the device, as indications of a low battery capacity. This can trigger an unnecessary low battery warning which, if it occurs during an rescue event may cause the device to switch off and/or may delay the delivery of shock therapy. Further information can be found at www.heartsine.com/recall